5.1 Normal Operating Conditions

5.1.1 The normal working environment for the sterilizer should meet the following conditions:

a) Ambient temperature: 5 ? to 40 ?;
b) Relative humidity: ? 80 % RH;
c) Atmospheric pressure: 70 kPa to 106 kPa;

5.1.2 Rated Operating Conditions of the Sterilizer

The rated operating conditions for each model of the sterilizer are shown in Table 2.

ሞዴልSupply VoltageFrequencyኃይል (kW)Current (A)
BCS-1380V50Hz1624
BCS-21624
ቢሲኤስ-31624
BCS-4.52741
ቢሲኤስ-52741
ቢሲኤስ-62741
BCS-82741
ቢሲኤስ-104365
ቢሲኤስ-155685
ቢሲኤስ-2081123
ቢሲኤስ-2681123
BCS-2890137
ቢሲኤስ-3090137
BCS-30-290137
ቢሲኤስ-35100152
BCS-35-2100152
BCS-40-2100152
ቢሲኤስ-48100152
BCS-52150228
BCS-55-2150228
BCS-57-2150228
BCS-63-2150228
BCS-70-2155236
BCS-78-2165251
BCS-86-2165251
BCS-101-2200304
BCS-120-2200304
  • The power supply socket is a three-phase 380V, three-hole socket with a grounding wire.
  • The power plug is a three-phase 380V, three-prong plug with a grounding wire.
  • A yellow-green outer wire is used to connect to the protective conductor terminal.

5.2 Storage Environment of the 12m3 eto sterilizer

The sterilizer should be stored in an indoor environment with temperatures between -20? and +55?, relative humidity ? 80% RH, free from corrosive gases, and well-ventilated.
ማስታወሻ፡- Before storage, all water in the circulating water system must be drained to keep the system in a dry state.

5.3 Transportation Conditions of the 12m3 eto sterilizer

The sterilizer should be transported unpackaged, and when hoisting the sterilizer, it should be lifted using the device’s own lifting rings. During transport by vehicle, it should be secured as per regulations.

5.4 Disposal of Decommissioned 12m3 Sterilizers

After decommissioning, the sterilizer should be disposed of by a qualified recycling company.