Protection mechanism and explanation OF ETO Sterilizer System
- ETO sterilizer system has six different levels of operation permissions. Personnel with different permissions can operate different contents to prevent operation confusion, or equipment failure or safety accident caused by operation by personnel without professional training
2. By default, the system will perform the leak detection test before dosing of each batch. If the leak detection test fails, the system will automatically stop the cycle and no longer continue, and the alarm will prompt

3. 如果設備曝光階段櫃體漏水, ETO 室 System 會自動跳到清潔階段,確保 EO 唔會滲入車間,造成安全事故
4. 當設備運行中,如果想中途停止循環,系統會根據唔同嘅運行條件自動跳轉到安全嘅處理過程。例如,喺加藥之後,個系統需要強制清潔嚟重設成個過程。
5. 如果係統配備電子秤,只要設備通電,就可以隨時檢查 EO 氣缸重量有否異常變化,以便及時通知設備維修人員潛在安全隱患
6. The ETO chamber System has the function of ultra-high pressure detection and protection. If the system detects that the pressure in the cabinet is greater than the maximum allowable pressure, whether the equipment is in operation or not, the vacuum pump must be forced to start to reduce the pressure in the cabinet to prevent the gas in the cabinet from leaking into the workshop, or the cabinet will be damaged by the high pressure in the cabinet
7.Each time the door is opened, the ETO chamber System will check whether the pressure in the cabinet is within the safe range, and whether the operation state in the cabinet is in the safe state, otherwise the door will be locked and cannot be opened
8. 為確保人身安全,設備可配備無害排氣系統。櫃門打開後,確保車間嘅空氣只可以進入櫃內,而唔可以由櫃排出到車間環境
如果設備已經清潔好,但係門仲未打開,系統會自動檢查卸貨超時。如果卸貨超時超過30分鐘,門將會被鎖住,無法打開,並且必須再次執行清潔程序。呢個功能係用嚟確保產品入面嘅 EO 唔會慢慢釋放到車間環境,對車間環境造成污染同個人健康受損
整個控制系統對各種設備同傳感器都有自我診斷功能,係用嚟防止傳感器故障時設備嘅異常運作,同時提示維修人員咩傳感器或設備出現故障,令維修人員可以快速搵到同解決問題
11. ETO 滅菌器 SYSTEM 有超過 100 個故障和警報信息,使操作員和維修人員能夠及時準確地知道設備運行情況,使維修人員能夠快速找到和解決問題
12. 下載工藝參數後,如果有不合理參數,會有下載失敗提示,防止參數不合理時整批運行失敗
13. 下載工藝參數後,如果有不合理參數,會有下載失敗提示,防止參數不合理時整批運行失敗
14. 如果外部電源斷電, UPS 會繼續為系統供電,設備會處理停機狀態,確保設備停電後唔會漏電,並確保安全緩衝時間至少為 15 分鐘。設備通過正常運行過程停機後,當外部電源恢復時,設備仍然可以繼續運行
15. 當系統運行期間 IPC 出現故障時,可以直接重新啟動 IPC ,軟件可以自動連接原始記錄
16. 系統具有詳細審計跟蹤功能,可以記錄幾乎所有操作動作,方便後期事故分析和故障預防
聯繫
電子郵件:hayley@hzbocon.com
本地站點:中國浙江省杭州市下沙區蔡通中新1202室