{"id":2425,"date":"2024-12-17T10:45:20","date_gmt":"2024-12-17T02:45:20","guid":{"rendered":"https:\/\/zjbocon.com\/?p=2425"},"modified":"2026-02-09T09:58:51","modified_gmt":"2026-02-09T01:58:51","slug":"eo-sterilization-equipment-and-process-characteristic-patterns-and-effect-analysis","status":"publish","type":"post","link":"https:\/\/zjbocon.com\/?p=2425","title":{"rendered":"EO sterilization equipment and process characteristic patterns and effect analysis"},"content":{"rendered":"\n<div class=\"wp-block-uagb-image uagb-block-de5b384b wp-block-uagb-image--layout-default wp-block-uagb-image--effect-static wp-block-uagb-image--align-none\"><figure class=\"wp-block-uagb-image__figure\"><img loading=\"lazy\" decoding=\"async\" src=\"https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp\" alt=\"\" class=\"uag-image-8473\" width=\"1024\" height=\"768\" title=\"ethylene-oxide-sterilization-68\" role=\"img\" \><\/figure><\/div>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Start cycle phase(Select parameters\/sterilization recipe)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"2\">1<\/td><td rowspan=\"2\">Choosing the wrong recipe<\/td><td rowspan=\"2\">Lack of care in choosing the formulation<\/td><td>Invalid cycle. Affects product sterility (re-sterilization)<\/td><td>1. Different recipe names<\/td><td>6<\/td><td>5<\/td><td>10<\/td><td>300<\/td><td>1. Work instructions 2. Careful inspection before the cycle starts 3. Operator training<\/td><td>2<\/td><td>5<\/td><td>4<\/td><td>40<\/td><td>1. Work instructions 2. Careful inspection before starting the cycle 3. Operator training<\/td><td>user<\/td><\/tr><tr><td>Loss of product\/packaging functionality and integrity due to unvalidated cycle parameters<\/td><td>1. Different recipe names<\/td><td>6<\/td><td>4<\/td><td>10<\/td><td>240<\/td><td>1. Work instructions 2. Careful inspection before the cycle starts 3. Operator training<\/td><td>2<\/td><td>4<\/td><td>4<\/td><td>32<\/td><td>1. Work instructions 2. Careful inspection before starting the cycle 3. Operator training<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Warm-up phase(Cabinet heating)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"3\">1<\/td><td rowspan=\"3\">The box temperature is lower than the lower limit value<\/td><td>Hot water circulation pump failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation<\/td><td>5<\/td><td>3<\/td><td>2<\/td><td>30<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Electric heating tube failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td><br>1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>2<\/td><td>The temperature inside the cabinet is too high<\/td><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td><br>1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Insulation stage(Maintain cabinet temperature)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"3\">1<\/td><td rowspan=\"3\">The box temperature is lower than the lower limit value<\/td><td>Hot water circulation pump failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm<\/td><td>5<\/td><td>3<\/td><td>2<\/td><td>30<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Electric heating tube failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>2<\/td><td>The temperature inside the cabinet is too high<\/td><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>3<\/td><td>Lack of time<\/td><td>Programming Errors<\/td><td>Repeat sterilization<\/td><td>1. Automatic cycle inspection<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td><\/td><td><\/td><td><\/td><td><\/td><td><\/td><td>none<\/td><td><\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Pre-vacuum stage(Removing air from the chamber allows the process to proceed below the EO flammability limit)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"3\">1<\/td><td rowspan=\"3\">Not enough air removed<\/td><td>Low vacuum pump efficiency<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. Pre-vacuum time over limit alarm 2. Vacuum pump overload alarm 3. Automatically detect water flow and shut down the waterless pump<\/td><td>5<\/td><td>3<\/td><td>2<\/td><td>30<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Wrong valve<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. Pre-vacuum time over limit alarm 2. Vacuum pump overload alarm<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Pressure sensor failure<\/td><td>Flammable Zone Hazard<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>2<\/td><td>Too much vacuum<\/td><td>Pressure sensor failure<\/td><td>Product and\/or packaging damage (classification)<\/td><td>1. Redundant pressure sensor 2. Pressure difference alarm is issued when the pressure difference is greater than 1kpa.<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Leak detection stage(Check the tightness of the box)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"3\">1<\/td><td rowspan=\"3\">Cabinet leakage<\/td><td>Gasket leakage<\/td><td>No impact<\/td><td>1. Leak test failure alarm 2. Process ends<\/td><td>10<\/td><td>5<\/td><td>2<\/td><td>100<\/td><td>1. Preventive maintenance<\/td><td>3<\/td><td>5<\/td><td>2<\/td><td>30<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Valve leakage<\/td><td>No impact<\/td><td>1. Leak test failure alarm<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Circulation system air duct leakage<\/td><td>No impact<\/td><td>1. Leak test failure alarm<\/td><td>4<\/td><td>3<\/td><td>2<\/td><td>24<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Nitrogen replacement stage(Fill the chamber with an inert gas to replace oxygen)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"3\">1<\/td><td rowspan=\"3\">Not enough nitrogen entering the chamber<\/td><td>Insufficient nitrogen<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>End of process<\/td><td>3<\/td><td>3<\/td><td>2<\/td><td>18<\/td><td><\/td><td><\/td><td><\/td><td><\/td><td><\/td><td>N\/A<\/td><td><\/td><\/tr><tr><td>Faulty valve<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>N\/A<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Pressure sensor failure<\/td><td>Flammable Zone Hazard<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td rowspan=\"2\">2<\/td><td rowspan=\"2\">Too much nitrogen in the cabinet door<\/td><td>Faulty valve<\/td><td>EO concentration is lower than the validated value because during EO injection, if controlled by pressure (resterilization), additional N2 is injected<\/td><td>1. Indoor high voltage alarm 2. End of process<\/td><td>6<\/td><td>5<\/td><td>4<\/td><td>120<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>5<\/td><td>4<\/td><td>40<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Pressure transmitter failure<\/td><td>Too much EO gas injected. High EO residue. Extended degassing time me)<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>3<\/td><td>The cabinet temperature is higher than the upper limit t<\/td><td><\/td><td>Impact on product function (sorting)<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td rowspan=\"4\">4<\/td><td rowspan=\"4\">The box temperature is lower than the lower limit value<\/td><td>Hot water circulation pump failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Electric heating tube failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>Low carburetor temperature<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Humidification stage(Increasing the humidity of the load by injecting steam)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"4\">1<\/td><td rowspan=\"4\">Insufficient steam<\/td><td>Steam generator failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Alarm when steam pressure is lower than the limit<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Valve failure<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1.Humidification frequency limit alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Pressure transmitter\/humidity sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Redundant pressure\/humidity sensor 2. Alarm when pressure\/humidity difference exceeds limit<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>Filter clogged<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1.Humidification frequency limit alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td rowspan=\"2\">2<\/td><td rowspan=\"2\">Too much steam injection<\/td><td>Faulty valve<\/td><td>Excessive steam injection, wet cartons (sorting)<\/td><td>1. Indoor high voltage alarm 2. End of process<\/td><td>6<\/td><td>5<\/td><td>4<\/td><td>120<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>5<\/td><td>4<\/td><td>40<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Pressure transmitter\/humidity sensor failure<\/td><td>Excessive steam injection, wet cartons (sorting)<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>3<\/td><td>Temperature is too high<\/td><td>Temperature sensor failure<\/td><td>Impact on product functions (ranking)<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td rowspan=\"3\">4<\/td><td rowspan=\"3\">The box temperature is lower than the lower limit value<\/td><td>Hot water circulation pump failure<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Electric heating tube failure<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. Heating timeout alarm 2. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Humidification dwell phase(Balanced humidity)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td>1<\/td><td>Lack of time<\/td><td>time out<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Automatic cycle inspection<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance<\/td><td>1<\/td><td>5<\/td><td>2<\/td><td>10<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Nitrogen cushion stage(Fill the chamber with inert gas to dilute the oxygen)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"3\">1<\/td><td rowspan=\"3\">Not enough nitrogen entering the cabinet<\/td><td>Lack of nitrogen<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. Check N2 pressure before starting the cycle 2. Alarm when N2 pressure is lower than the limit<\/td><td>3<\/td><td>3<\/td><td>2<\/td><td>18<\/td><td><\/td><td><\/td><td><\/td><td><\/td><td>0<\/td><td><\/td><td><\/td><\/tr><tr><td>Valve failure<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>none<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>user<\/td><td>1. Preventive maintenance<\/td><\/tr><tr><td>Pressure sensor failure<\/td><td>Flammable Zone Hazard<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>user<\/td><td>1. Preventive maintenance<\/td><\/tr><tr><td rowspan=\"2\">2<\/td><td rowspan=\"2\">Too much nitrogen injected into the cabinet<\/td><td>Faulty valve<\/td><td>EO concentration is lower than the validated value because during EO injection, if controlled by pressure (resterilization), additional N2 is injected<\/td><td>1. Indoor high voltage alarm 2. End of process<\/td><td>6<\/td><td>5<\/td><td>4<\/td><td>120<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>5<\/td><td>4<\/td><td>40<\/td><td>user<\/td><td>1. Preventive maintenance<\/td><\/tr><tr><td>Pressure transmitter failure<\/td><td>Too much EO gas injected. High EO residue. (Extend degassing time)<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>user<\/td><td>1. Preventive maintenance 2. Calibration<\/td><\/tr><tr><td>3<\/td><td>The temperature inside the cabinet exceeds the upper limit<\/td><td>Temperature sensor failure<\/td><td>Impact on product function (sorting)<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>user<\/td><td>1. Preventive maintenance 2. Calibration<\/td><\/tr><tr><td rowspan=\"4\">4<\/td><td rowspan=\"4\">The box temperature is lower than the lower limit value<\/td><td>Hot water circulation pump failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>user<\/td><td>1. Preventive maintenance<\/td><\/tr><tr><td>Electric heating tube failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>user<\/td><td>1. Preventive maintenance<\/td><\/tr><tr><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>user<\/td><td>1. Preventive maintenance 2. Calibration<\/td><\/tr><tr><td>Low carburetor temperature<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>user<\/td><td>1. Preventive maintenance 2. Calibration<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">EO injection stage(Fill the cabinet with EO gas)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"4\">1<\/td><td rowspan=\"4\">Not enough ethylene oxide is entering the cabinet<\/td><td>Missing EO<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1.EO cylinder hole alarm 2.EO injection timeout alarm<\/td><td>10<\/td><td>5<\/td><td>2<\/td><td>100<\/td><td>1. Operator training 2. EO weight check before cycle start<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Operator training 2. EO weight check before cycle start<\/td><td>user<\/td><\/tr><tr><td>Faulty valve<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1.EO injection timeout alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>EO pipeline leakage<\/td><td>Batches that exceed the qualification period may be sterilized again<\/td><td>none<\/td><td>6<\/td><td>5<\/td><td>3<\/td><td>90<\/td><td>1. Preventive maintenance 2. LEL EO detector alarm<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. LEL EO detector alarm<\/td><td>user<\/td><\/tr><tr><td>Pressure sensor failure<\/td><td>EO concentration is lower than the validation value (re-sterilization)<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td rowspan=\"2\">2<\/td><td rowspan=\"2\">Liquid EO enters the cabinet<\/td><td>Low carburetor temperature<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. The carburetor temperature is lower than the limit alarm<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Not enough water in the carburetor due to a faulty valve<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. The carburetor water level is lower than the limit alarm<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td rowspan=\"2\">3<\/td><td rowspan=\"2\">Too much ethylene oxide input in the cabinet<\/td><td>Faulty valve<\/td><td>Longer quarantine, requiring additional testing<\/td><td>1. Indoor high voltage alarm<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance 2. EO residual inspection<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>1. Preventive maintenance 2. EO residual inspection<\/td><td>user<\/td><\/tr><tr><td>Pressure sensor failure<\/td><td>Longer quarantine, requiring additional testing<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>4<\/td><td>The temperature inside the cabinet exceeds the upper limit<\/td><td>Temperature sensor failure<\/td><td>Impact on product function (sorting)<\/td><td>1. Redundant temperature sensor 2. Chamber high temperature alarm 3. Temperature difference exceeds limit alarm<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td rowspan=\"4\">5<\/td><td rowspan=\"4\">The box temperature is lower than the lower limit value<\/td><td>Hot water circulation pump failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Electric heating tube failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>Low carburetor temperature<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Nitrogen cap stage(Add N2 to cover EO)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"3\">1<\/td><td rowspan=\"3\">Not enough nitrogen entering the cabinet<\/td><td>Lack of nitrogen<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. Check N2 pressure before starting the cycle 2. Alarm when N2 pressure is lower than the limit<\/td><td>3<\/td><td>3<\/td><td>2<\/td><td>18<\/td><td><\/td><td><\/td><td><\/td><td><\/td><td><\/td><td>N\/A<\/td><td><\/td><\/tr><tr><td>Valve failure<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>none<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Pressure sensor failure<\/td><td>Flammable Zone Hazard<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>2<\/td><td>Too much nitrogen injected into the cabinet<\/td><td>Faulty valve<\/td><td>The cabin is under positive pressure. This cycle will end.<\/td><td>1. Indoor high voltage alarm 2. End of process<\/td><td>6<\/td><td>5<\/td><td>4<\/td><td>120<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>5<\/td><td>4<\/td><td>40<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>3<\/td><td>The temperature inside the cabinet exceeds the upper limit<\/td><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Redundant temperature sensor 2. Chamber high temperature alarm 3. Temperature difference exceeds limit alarm<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td rowspan=\"4\">4<\/td><td rowspan=\"4\">The box temperature is lower than the lower limit value<\/td><td>Hot water circulation pump failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Electric heating tube failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>Low carburetor temperature<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">EO exposure stage(Disinfect the product)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td>1<\/td><td>Lack of time<\/td><td>Programming Errors<\/td><td>Repeat sterilization<\/td><td>1. Automatic cycle inspection<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance<\/td><td>1<\/td><td>5<\/td><td>2<\/td><td>10<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td rowspan=\"3\">2<\/td><td rowspan=\"3\">The box temperature is lower than the lower limit t<\/td><td>Hot water circulation pump failure<\/td><td>The cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>Heating timeout 2. Circulating water pump overload alarm 3. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Electric heating tube failure<\/td><td>The cycle is locked in one phase until it is resumed and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.<\/td><td>1. Heating timeout alarm 2. Automatic heating time compensation<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm<\/td><td>6<\/td><td>5<\/td><td>2<\/td><td>60<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>5<\/td><td>2<\/td><td>20<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>3<\/td><td>The temperature inside the cabinet exceeds the upper limit<\/td><td>Temperature sensor failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Redundant temperature sensor 2. Chamber high temperature alarm 3. Temperature alarm<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td rowspan=\"2\">4<\/td><td rowspan=\"2\">Not getting a sufficiently homogeneous EO mixture<\/td><td>Circulation fan failure<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1. Low speed alarm<\/td><td>6<\/td><td>5<\/td><td>3<\/td><td>90<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>5<\/td><td>3<\/td><td>30<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Product positioning very close to the cavity wall<\/td><td>Affects product sterility (re-sterilization)<\/td><td>1.Guardrail is used to prevent the product from contacting the cavity wall<\/td><td>2<\/td><td>5<\/td><td>3<\/td><td>30<\/td><td><\/td><td><\/td><td><\/td><td><\/td><td>0<\/td><td>N\/A<\/td><td><\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">EO Removal Phase(Create vacuum\/remove sterilization gas)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"3\">1<\/td><td rowspan=\"3\">Not purging enough gas<\/td><td>Inefficient vacuum pump<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1.EO removal timeout alarm 2.Vacuum pump overload alarm 3.Automatically detect water flow and shut down the waterless pump<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Faulty valve<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1.EO removal timeout alarm 2.Vacuum pump overload alarm<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Pressure sensor failure<\/td><td>Longer isolation time and additional testing required due to less EO gas being removed<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>2<\/td><td>Exhaust too much gas<\/td><td>Pressure sensor failure<\/td><td>Product and\/or packaging damage (classification)<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Nitrogen purge stage(Ensure that the indoor environment is not flammable)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"4\">1<\/td><td rowspan=\"4\">Not enough nitrogen entering the chamber<\/td><td>Lack of nitrogen<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. Check N2 pressure before starting the cycle 2. Alarm when N2 pressure is lower than the limit<\/td><td>3<\/td><td>3<\/td><td>2<\/td><td>18<\/td><td><\/td><td><\/td><td><\/td><td><\/td><td><\/td><td>N\/A<\/td><td><\/td><\/tr><tr><td>Faulty valve<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>none<\/td><td>6<\/td><td>3<\/td><td>3<\/td><td>54<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>3<\/td><td>18<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td rowspan=\"2\">Pressure sensor failure<\/td><td>Flammable Zone Hazard<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Longer quarantines require additional testing<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Air cleaning stage(Create vacuum\/remove residual sterilization gas, eliminate residual EO)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td rowspan=\"3\">1<\/td><td rowspan=\"3\">Not purging enough gas<\/td><td>Inefficient vacuum pump<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1.EO removal timeout alarm 2.Vacuum pump overload alarm 3.Automatically detect water flow and shut down the waterless pump<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Faulty valve<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1.EO removal timeout alarm 2.Vacuum pump overload alarm<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Pressure sensor failure<\/td><td>Longer isolation time and additional testing required due to less EO gas being removed<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td>2<\/td><td>Exhaust too much gas<\/td><td>Pressure sensor failure<\/td><td>Product and\/or packaging damage (classification)<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>4<\/td><td>2<\/td><td>48<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><tr><td rowspan=\"3\">3<\/td><td rowspan=\"3\">Insufficient air in the cabin<\/td><td>Faulty valve<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. Gas injection time over limit alarm<\/td><td>6<\/td><td>4<\/td><td>3<\/td><td>72<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>4<\/td><td>3<\/td><td>24<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Filter clogged<\/td><td>The cycle is locked in one phase until it is restored and restarted after a standby period<\/td><td>1. Gas injection time over limit alarm<\/td><td>6<\/td><td>4<\/td><td>3<\/td><td>72<\/td><td>1. Preventive maintenance<\/td><td>2<\/td><td>4<\/td><td>3<\/td><td>24<\/td><td>1. Preventive maintenance<\/td><td>user<\/td><\/tr><tr><td>Pressure sensor failure<\/td><td>Longer quarantines require additional testing<\/td><td>1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit<\/td><td>6<\/td><td>3<\/td><td>2<\/td><td>36<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>2<\/td><td>4<\/td><td>2<\/td><td>16<\/td><td>1. Preventive maintenance 2. Calibration<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td colspan=\"16\">Removal stage(Remove residue from the product)<\/td><\/tr><tr><td rowspan=\"2\">Serial number<\/td><td colspan=\"3\">Potentially risky content<\/td><td colspan=\"5\">Risk assessment under existing process conditions<\/td><td colspan=\"5\">Risk assessment of the results after taking measures<\/td><td rowspan=\"2\">Recommended Actions<\/td><td rowspan=\"2\">Responsibilities and current status<\/td><\/tr><tr><td>Fault name<\/td><td>Cause<\/td><td>Failure Impact<\/td><td>Control Now<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><td>Take action<\/td><td>event<\/td><td>Severity<\/td><td>examine<\/td><td>Risk Prioritization<\/td><\/tr><tr><td>1<\/td><td>Aeration time is less than the specified time<\/td><td>Programming Errors<\/td><td>Repeat sterilization<\/td><td>1. Automatic cycle inspection<\/td><td>2<\/td><td>3<\/td><td>2<\/td><td>12<\/td><td>1. Restart your computer regularly<\/td><td>1<\/td><td>3<\/td><td>2<\/td><td>6<\/td><td>1. Restart your computer regularly<\/td><td>user<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\" id=\"h-if-you-want-to-know-more-about-ethylene-oxide-sterilization\">IF YOU WANT TO KNOW MORE About Ethylene Oxide Sterilization<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Phone:&nbsp;<a href=\"https:\/\/wa.me\/8619975258603\">+8619975258603<\/a><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Email:&nbsp;<a href=\"mailto:hayley@hzbocon.com\">hayley@hzbocon.com<\/a><\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Local Site: Room 1202, Caitong Zhongxin, Xiasha District, Hangzhou City, Zhejiang Province, China<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Website:&nbsp;<a href=\"https:\/\/www.hzbocon.com\/\">hzbocon.com<\/a><a href=\"https:\/\/zjbocon.com\/\">&nbsp;zjbocon.com<\/a><\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><\/p>\n","protected":false},"excerpt":{"rendered":"<p>Start cycle phase(Select parameters\/sterilization recipe) Serial number Potentially risky content Risk assessment under existing process conditions Risk assessment of the [&hellip;]<\/p>\n","protected":false},"author":234312599,"featured_media":8473,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_uag_custom_page_level_css":"","site-sidebar-layout":"default","site-content-layout":"","ast-site-content-layout":"","site-content-style":"default","site-sidebar-style":"default","ast-global-header-display":"","ast-banner-title-visibility":"","ast-main-header-display":"","ast-hfb-above-header-display":"","ast-hfb-below-header-display":"","ast-hfb-mobile-header-display":"","site-post-title":"","ast-breadcrumbs-content":"","ast-featured-img":"","footer-sml-layout":"","ast-disable-related-posts":"","theme-transparent-header-meta":"","adv-header-id-meta":"","stick-header-meta":"","header-above-stick-meta":"","header-main-stick-meta":"","header-below-stick-meta":"","astra-migrate-meta-layouts":"set","ast-page-background-enabled":"default","ast-page-background-meta":{"desktop":{"background-color":"","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""},"tablet":{"background-color":"","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""},"mobile":{"background-color":"","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""}},"ast-content-background-meta":{"desktop":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""},"tablet":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""},"mobile":{"background-color":"var(--ast-global-color-5)","background-image":"","background-repeat":"repeat","background-position":"center center","background-size":"auto","background-attachment":"scroll","background-type":"","background-media":"","overlay-type":"","overlay-color":"","overlay-opacity":"","overlay-gradient":""}},"_jetpack_newsletter_access":"","_jetpack_dont_email_post_to_subs":false,"_jetpack_newsletter_tier_id":0,"_jetpack_memberships_contains_paywalled_content":false,"_jetpack_feature_clip_id":0,"_jetpack_memberships_contains_paid_content":false,"footnotes":"","jetpack_publicize_message":"","jetpack_publicize_feature_enabled":true,"jetpack_social_post_already_shared":false,"jetpack_social_options":{"image_generator_settings":{"template":"highway","enabled":false},"version":2},"jetpack_post_was_ever_published":false},"categories":[133124],"tags":[],"class_list":["post-2425","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-eto-sterilizer"],"yoast_head":"<!-- This site is optimized with the Yoast SEO Premium plugin v27.2 (Yoast SEO v28.0) - https:\/\/yoast.com\/product\/yoast-seo-premium-wordpress\/ -->\n<title>EO sterilization equipment and process characteristic patterns and effect analysis - ETO STERILIZER<\/title>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \>\n<link rel=\"canonical\" href=\"https:\/\/zjbocon.com\/?p=2425\" \>\n<meta property=\"og:locale\" content=\"en_US\" \>\n<meta property=\"og:type\" content=\"article\" \>\n<meta property=\"og:title\" content=\"EO sterilization equipment and process characteristic patterns effect analysis\" \>\n<meta property=\"og:description\" content=\"Start cycle phase(Select parameters\>\n<meta property=\"og:url\" content=\"https:\/\/zjbocon.com\/?p=2425\" \>\n<meta property=\"og:site_name\" content=\"ETO STERILIZER\" \>\n<meta property=\"article:publisher\" content=\"https:\/\/www.facebook.com\/profile.php?id=100088482339878\" \>\n<meta property=\"article:author\" content=\"https:\/\/www.facebook.com\/profile.php?id=100088482339878\" \>\n<meta property=\"article:published_time\" content=\"2024-12-17T02:45:20+00:00\" \>\n<meta property=\"article:modified_time\" content=\"2026-02-09T01:58:51+00:00\" \>\n<meta property=\"og:image\" content=\"https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp\" \>\n\t<meta property=\"og:image:width\" content=\"384\" \>\n\t<meta property=\"og:image:height\" content=\"512\" \>\n\t<meta property=\"og:image:type\" content=\"image\/webp\" \>\n<meta name=\"author\" content=\"Bocon Technical Team\" \>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \>\n<meta name=\"twitter:creator\" content=\"@BoconAnd\" \>\n<meta name=\"twitter:site\" content=\"@BoconAnd\" \>\n<meta name=\"twitter:label1\" content=\"Written by\" \>\n\t<meta name=\"twitter:data1\" content=\"Bocon Technical Team\" \>\n\t<meta name=\"twitter:label2\" content=\"Est. reading time\" \>\n\t<meta name=\"twitter:data2\" content=\"23 minutes\" \>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425#article\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425\"},\"author\":{\"name\":\"Bocon Technical Team\",\"@id\":\"https:\\\/\\\/zjbocon.com\\\/#\\\/schema\\\/person\\\/f12be028b8ae9843a4396dcc8012b425\"},\"headline\":\"EO sterilization equipment and process characteristic patterns and effect analysis\",\"datePublished\":\"2024-12-17T02:45:20+00:00\",\"dateModified\":\"2026-02-09T01:58:51+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425\"},\"wordCount\":3962,\"commentCount\":0,\"publisher\":{\"@id\":\"https:\\\/\\\/zjbocon.com\\\/#organization\"},\"image\":{\"@id\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/zjbocon.com\\\/wp-content\\\/uploads\\\/2025\\\/09\\\/ethylene-oxide-sterilization-68.webp\",\"articleSection\":[\"ETO STERILIZER\"],\"inLanguage\":\"mrj\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\\\/\\\/zjbocon.com\\\/?p=2425#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425\",\"url\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425\",\"name\":\"EO sterilization equipment and process characteristic patterns and effect analysis - ETO STERILIZER\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/zjbocon.com\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/zjbocon.com\\\/wp-content\\\/uploads\\\/2025\\\/09\\\/ethylene-oxide-sterilization-68.webp\",\"datePublished\":\"2024-12-17T02:45:20+00:00\",\"dateModified\":\"2026-02-09T01:58:51+00:00\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425#breadcrumb\"},\"inLanguage\":\"mrj\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/zjbocon.com\\\/?p=2425\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"mrj\",\"@id\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425#primaryimage\",\"url\":\"https:\\\/\\\/zjbocon.com\\\/wp-content\\\/uploads\\\/2025\\\/09\\\/ethylene-oxide-sterilization-68.webp\",\"contentUrl\":\"https:\\\/\\\/zjbocon.com\\\/wp-content\\\/uploads\\\/2025\\\/09\\\/ethylene-oxide-sterilization-68.webp\",\"width\":384,\"height\":512},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/zjbocon.com\\\/?p=2425#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/zjbocon.com\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"EO sterilization equipment and process characteristic patterns and effect analysis\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/zjbocon.com\\\/#website\",\"url\":\"https:\\\/\\\/zjbocon.com\\\/\",\"name\":\"ETO STERILIZER\",\"description\":\"One-stop Solution Supplier of Ethylene Oxide Sterilization\",\"publisher\":{\"@id\":\"https:\\\/\\\/zjbocon.com\\\/#organization\"},\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/zjbocon.com\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"mrj\"},{\"@type\":\"Organization\",\"@id\":\"https:\\\/\\\/zjbocon.com\\\/#organization\",\"name\":\"ETO STERILIZER\",\"url\":\"https:\\\/\\\/zjbocon.com\\\/\",\"logo\":{\"@type\":\"ImageObject\",\"inLanguage\":\"mrj\",\"@id\":\"https:\\\/\\\/zjbocon.com\\\/#\\\/schema\\\/logo\\\/image\\\/\",\"url\":\"https:\\\/\\\/zjbocon.com\\\/wp-content\\\/uploads\\\/2024\\\/10\\\/logo-1.webp\",\"contentUrl\":\"https:\\\/\\\/zjbocon.com\\\/wp-content\\\/uploads\\\/2024\\\/10\\\/logo-1.webp\",\"width\":180,\"height\":60,\"caption\":\"ETO STERILIZER\"},\"image\":{\"@id\":\"https:\\\/\\\/zjbocon.com\\\/#\\\/schema\\\/logo\\\/image\\\/\"},\"sameAs\":[\"https:\\\/\\\/www.facebook.com\\\/profile.php?id=100088482339878\",\"https:\\\/\\\/x.com\\\/BoconAnd\",\"https:\\\/\\\/www.linkedin.com\\\/in\\\/--06a221139\\\/\"]},{\"@type\":\"Person\",\"@id\":\"https:\\\/\\\/zjbocon.com\\\/#\\\/schema\\\/person\\\/f12be028b8ae9843a4396dcc8012b425\",\"name\":\"Bocon Technical Team\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"mrj\",\"@id\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d7a825658be227f26b237e519f8541e05aabc9ff27b8eec4544a233a48c53be0?s=96&d=identicon&r=g\",\"url\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d7a825658be227f26b237e519f8541e05aabc9ff27b8eec4544a233a48c53be0?s=96&d=identicon&r=g\",\"contentUrl\":\"https:\\\/\\\/secure.gravatar.com\\\/avatar\\\/d7a825658be227f26b237e519f8541e05aabc9ff27b8eec4544a233a48c53be0?s=96&d=identicon&r=g\",\"caption\":\"Bocon Technical Team\"},\"sameAs\":[\"https:\\\/\\\/zjbocon.com\\\/\",\"https:\\\/\\\/www.facebook.com\\\/profile.php?id=100088482339878\",\"https:\\\/\\\/www.linkedin.com\\\/in\\\/\\\/hayley--968190269\\\/\"],\"url\":\"https:\\\/\\\/zjbocon.com\\\/?author=234312599\"}]}<\/script>\n<!-- \/ Yoast SEO Premium plugin. -->","yoast_head_json":{"title":"EO sterilization equipment and process characteristic patterns and effect analysis - ETO STERILIZER","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/zjbocon.com\/?p=2425","og_locale":"en_US","og_type":"article","og_title":"EO sterilization equipment and process characteristic patterns and effect analysis","og_description":"Start cycle phase(Select parameters\/sterilization recipe) Serial number Potentially risky content Risk assessment under existing process conditions Risk assessment of the [&hellip;]","og_url":"https:\/\/zjbocon.com\/?p=2425","og_site_name":"ETO STERILIZER","article_publisher":"https:\/\/www.facebook.com\/profile.php?id=100088482339878","article_author":"https:\/\/www.facebook.com\/profile.php?id=100088482339878","article_published_time":"2024-12-17T02:45:20+00:00","article_modified_time":"2026-02-09T01:58:51+00:00","og_image":[{"width":384,"height":512,"url":"https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp","type":"image\/webp"}],"author":"Bocon Technical Team","twitter_card":"summary_large_image","twitter_creator":"@BoconAnd","twitter_site":"@BoconAnd","twitter_misc":{"Written by":"Bocon Technical Team","Est. reading time":"23 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/zjbocon.com\/?p=2425#article","isPartOf":{"@id":"https:\/\/zjbocon.com\/?p=2425"},"author":{"name":"Bocon Technical Team","@id":"https:\/\/zjbocon.com\/#\/schema\/person\/f12be028b8ae9843a4396dcc8012b425"},"headline":"EO sterilization equipment and process characteristic patterns and effect analysis","datePublished":"2024-12-17T02:45:20+00:00","dateModified":"2026-02-09T01:58:51+00:00","mainEntityOfPage":{"@id":"https:\/\/zjbocon.com\/?p=2425"},"wordCount":3962,"commentCount":0,"publisher":{"@id":"https:\/\/zjbocon.com\/#organization"},"image":{"@id":"https:\/\/zjbocon.com\/?p=2425#primaryimage"},"thumbnailUrl":"https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp","articleSection":["ETO STERILIZER"],"inLanguage":"mrj","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/zjbocon.com\/?p=2425#respond"]}]},{"@type":"WebPage","@id":"https:\/\/zjbocon.com\/?p=2425","url":"https:\/\/zjbocon.com\/?p=2425","name":"EO sterilization equipment and process characteristic patterns and effect analysis - ETO STERILIZER","isPartOf":{"@id":"https:\/\/zjbocon.com\/#website"},"primaryImageOfPage":{"@id":"https:\/\/zjbocon.com\/?p=2425#primaryimage"},"image":{"@id":"https:\/\/zjbocon.com\/?p=2425#primaryimage"},"thumbnailUrl":"https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp","datePublished":"2024-12-17T02:45:20+00:00","dateModified":"2026-02-09T01:58:51+00:00","breadcrumb":{"@id":"https:\/\/zjbocon.com\/?p=2425#breadcrumb"},"inLanguage":"mrj","potentialAction":[{"@type":"ReadAction","target":["https:\/\/zjbocon.com\/?p=2425"]}]},{"@type":"ImageObject","inLanguage":"mrj","@id":"https:\/\/zjbocon.com\/?p=2425#primaryimage","url":"https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp","contentUrl":"https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp","width":384,"height":512},{"@type":"BreadcrumbList","@id":"https:\/\/zjbocon.com\/?p=2425#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/zjbocon.com\/"},{"@type":"ListItem","position":2,"name":"EO sterilization equipment and process characteristic patterns and effect analysis"}]},{"@type":"WebSite","@id":"https:\/\/zjbocon.com\/#website","url":"https:\/\/zjbocon.com\/","name":"ETO STERILIZER","description":"One-stop Solution Supplier of Ethylene Oxide Sterilization","publisher":{"@id":"https:\/\/zjbocon.com\/#organization"},"potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/zjbocon.com\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"mrj"},{"@type":"Organization","@id":"https:\/\/zjbocon.com\/#organization","name":"ETO STERILIZER","url":"https:\/\/zjbocon.com\/","logo":{"@type":"ImageObject","inLanguage":"mrj","@id":"https:\/\/zjbocon.com\/#\/schema\/logo\/image\/","url":"https:\/\/zjbocon.com\/wp-content\/uploads\/2024\/10\/logo-1.webp","contentUrl":"https:\/\/zjbocon.com\/wp-content\/uploads\/2024\/10\/logo-1.webp","width":180,"height":60,"caption":"ETO STERILIZER"},"image":{"@id":"https:\/\/zjbocon.com\/#\/schema\/logo\/image\/"},"sameAs":["https:\/\/www.facebook.com\/profile.php?id=100088482339878","https:\/\/x.com\/BoconAnd","https:\/\/www.linkedin.com\/in\/--06a221139\/"]},{"@type":"Person","@id":"https:\/\/zjbocon.com\/#\/schema\/person\/f12be028b8ae9843a4396dcc8012b425","name":"Bocon Technical Team","image":{"@type":"ImageObject","inLanguage":"mrj","@id":"https:\/\/secure.gravatar.com\/avatar\/d7a825658be227f26b237e519f8541e05aabc9ff27b8eec4544a233a48c53be0?s=96&d=identicon&r=g","url":"https:\/\/secure.gravatar.com\/avatar\/d7a825658be227f26b237e519f8541e05aabc9ff27b8eec4544a233a48c53be0?s=96&d=identicon&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/d7a825658be227f26b237e519f8541e05aabc9ff27b8eec4544a233a48c53be0?s=96&d=identicon&r=g","caption":"Bocon Technical Team"},"sameAs":["https:\/\/zjbocon.com\/","https:\/\/www.facebook.com\/profile.php?id=100088482339878","https:\/\/www.linkedin.com\/in\/\/hayley--968190269\/"],"url":"https:\/\/zjbocon.com\/?author=234312599"}]}},"jetpack_publicize_connections":[],"jetpack_featured_media_url":"https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp","uagb_featured_image_src":{"full":["https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp",384,512,false],"thumbnail":["https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68-150x150.webp",150,150,true],"medium":["https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68-225x300.webp",225,300,true],"medium_large":["https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp",384,512,false],"large":["https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp",384,512,false],"1536x1536":["https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp",384,512,false],"2048x2048":["https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp",384,512,false],"woocommerce_thumbnail":["https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp",384,512,false],"woocommerce_single":["https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68.webp",384,512,false],"woocommerce_gallery_thumbnail":["https:\/\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-68-100x100.webp",100,100,true]},"uagb_author_info":{"display_name":"Bocon Technical Team","author_link":"https:\/\/zjbocon.com\/?author=234312599"},"uagb_comment_info":18,"uagb_excerpt":"Start cycle phase(Select parameters\/sterilization recipe) Serial number Potentially risky content Risk assessment under existing process conditions Risk assessment of the [&hellip;]","jetpack_likes_enabled":true,"jetpack_sharing_enabled":true,"jetpack_shortlink":"https:\/\/wp.me\/pgZDV6-D7","jetpack-related-posts":[{"id":1950,"url":"https:\/\/zjbocon.com\/?p=1950","url_meta":{"origin":2425,"position":0},"title":"eto sterilization process","author":"Bocon Technical Team","date":"2024-10-23","format":false,"excerpt":"Eto Sterilization is conducted using a eto sterilizer located in a dedicated facility.","rel":"","context":"In &quot;ETO STERILIZER&quot;","block_context":{"text":"ETO STERILIZER","link":"https:\/\/zjbocon.com\/?cat=133124"},"img":{"alt_text":"eto sterilization process","src":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/10\/eto-sterilization-process-e1729654271899.webp?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/10\/eto-sterilization-process-e1729654271899.webp?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/10\/eto-sterilization-process-e1729654271899.webp?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/10\/eto-sterilization-process-e1729654271899.webp?resize=700%2C400&ssl=1 2x"},"classes":[]},{"id":2094,"url":"https:\/\/zjbocon.com\/?p=2094","url_meta":{"origin":2425,"position":1},"title":"ETO Sterilizer Manufacturer-HZBOCON","author":"Bocon Technical Team","date":"2024-11-04","format":false,"excerpt":"Hangzhou Bocon Mechanical and Electrical Equipment Co., Ltd., commonly referred to as HZBOCON, stands out as a dedicated Ethylene Oxide (EO\/ETO) sterilizer manufacturer. Established in 2006, HZBOCON has committed itself to the design, development, and manufacturing of EO sterilization equipment, catering to a wide array of industries that require effective\u2026","rel":"","context":"In &quot;ETO STERILIZER&quot;","block_context":{"text":"ETO STERILIZER","link":"https:\/\/zjbocon.com\/?cat=133124"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/eto-sterilizer-hzbocon-4.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/eto-sterilizer-hzbocon-4.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/eto-sterilizer-hzbocon-4.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/eto-sterilizer-hzbocon-4.png?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/eto-sterilizer-hzbocon-4.png?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":2032,"url":"https:\/\/zjbocon.com\/?p=2032","url_meta":{"origin":2425,"position":2},"title":"The advantage of Ethylene oxide sterilizer","author":"Bocon Technical Team","date":"2024-10-29","format":false,"excerpt":"Ethylene oxide sterilizer operate at low temperatures, making them ideal for sterilizing heat-sensitive medical devices and instruments. This characteristic allows for the preservation of the integrity and functionality of complex medical devices that might be damaged by heat.","rel":"","context":"In &quot;ETO STERILIZER&quot;","block_context":{"text":"ETO STERILIZER","link":"https:\/\/zjbocon.com\/?cat=133124"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/eto-sterilizer-hzbocon-4.png?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/eto-sterilizer-hzbocon-4.png?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/eto-sterilizer-hzbocon-4.png?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/eto-sterilizer-hzbocon-4.png?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/eto-sterilizer-hzbocon-4.png?resize=1050%2C600&ssl=1 3x"},"classes":[]},{"id":2224,"url":"https:\/\/zjbocon.com\/?p=2224","url_meta":{"origin":2425,"position":3},"title":"Factors Affecting the EO Gas Sterilization Process","author":"Bocon Technical Team","date":"2024-11-22","format":false,"excerpt":"Within a certain range of gas concentration, the higher the concentration, the better theEO Gas Sterilization effect; beyond 1000 mg\/L, the improvement in sterilization effect is not significant.","rel":"","context":"In &quot;ETO STERILIZER&quot;","block_context":{"text":"ETO STERILIZER","link":"https:\/\/zjbocon.com\/?cat=133124"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-10-scaled.webp?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-10-scaled.webp?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-10-scaled.webp?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-10-scaled.webp?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-10-scaled.webp?resize=1050%2C600&ssl=1 3x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-10-scaled.webp?resize=1400%2C800&ssl=1 4x"},"classes":[]},{"id":3093,"url":"https:\/\/zjbocon.com\/?p=3093","url_meta":{"origin":2425,"position":4},"title":"Introduction to Ethylene Oxide Sterilizers","author":"Bocon Technical Team","date":"2025-02-05","format":false,"excerpt":"Discover how ethylene oxide sterilizers work, their applications, benefits, and safety considerations. Learn why they are essential for medical and industrial sterilization.","rel":"","context":"In &quot;Company News&quot;","block_context":{"text":"Company News","link":"https:\/\/zjbocon.com\/?cat=133125"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2024\/11\/ethylene-oxide-sterilization-21.webp?resize=350%2C200","width":350,"height":200},"classes":[]},{"id":3190,"url":"https:\/\/zjbocon.com\/?p=3190","url_meta":{"origin":2425,"position":5},"title":"What Can Be Sterilized by Ethylene Oxide?","author":"Bocon Technical Team","date":"2025-02-12","format":false,"excerpt":"Ethylene Oxide (EO) gas sterilization is a highly effective method for sterilizing a wide range of items that are sensitive to heat, moisture, or radiation. ETO Sterile Medical Machines and ETO sterilizers are widely used in healthcare, pharmaceuticals, and industrial settings due to their ability to penetrate complex materials and\u2026","rel":"","context":"In &quot;Company News&quot;","block_context":{"text":"Company News","link":"https:\/\/zjbocon.com\/?cat=133125"},"img":{"alt_text":"","src":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-25.webp?resize=350%2C200&ssl=1","width":350,"height":200,"srcset":"https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-25.webp?resize=350%2C200&ssl=1 1x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-25.webp?resize=525%2C300&ssl=1 1.5x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-25.webp?resize=700%2C400&ssl=1 2x, https:\/\/i0.wp.com\/zjbocon.com\/wp-content\/uploads\/2025\/09\/ethylene-oxide-sterilization-25.webp?resize=1050%2C600&ssl=1 3x"},"classes":[]}],"_links":{"self":[{"href":"https:\/\/zjbocon.com\/index.php?rest_route=\/wp\/v2\/posts\/2425","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/zjbocon.com\/index.php?rest_route=\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/zjbocon.com\/index.php?rest_route=\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/zjbocon.com\/index.php?rest_route=\/wp\/v2\/users\/234312599"}],"replies":[{"embeddable":true,"href":"https:\/\/zjbocon.com\/index.php?rest_route=%2Fwp%2Fv2%2Fcomments&post=2425"}],"version-history":[{"count":13,"href":"https:\/\/zjbocon.com\/index.php?rest_route=\/wp\/v2\/posts\/2425\/revisions"}],"predecessor-version":[{"id":8866,"href":"https:\/\/zjbocon.com\/index.php?rest_route=\/wp\/v2\/posts\/2425\/revisions\/8866"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/zjbocon.com\/index.php?rest_route=\/wp\/v2\/media\/8473"}],"wp:attachment":[{"href":"https:\/\/zjbocon.com\/index.php?rest_route=%2Fwp%2Fv2%2Fmedia&parent=2425"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/zjbocon.com\/index.php?rest_route=%2Fwp%2Fv2%2Fcategories&post=2425"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/zjbocon.com\/index.php?rest_route=%2Fwp%2Fv2%2Ftags&post=2425"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}