BOCON Hot Air Eto Sterilizer Specifications
Compliance: EN 1422, CE, ISO 13485 Support: FAT, SAT, DQ, IQ, OQ, PQ
Suporte de Engenharia
Custom dimensions and automated conveyor systems are available upon request.
INQUIRE NOW →Bocon Hot Air Jacket ETO Sterilizer VS Traditional Water Jacket EO Sterilizer
| Feature / Specification | Bocon Hot Air Jacket System | Traditional Water Jacket System |
|---|---|---|
| Heating Efficiency | High: Uniformly blankets chamber and doors directly. | Lower: Indirect heat transfer via water medium. |
| Energy Consumption | Optimized: Only heats product load and air. | High: Significant energy wasted to maintain water temperature. |
| Heat-up Time (from 21°C) | Rapid: Less than 3 hours. | Slow: Requires heating large volumes of water first. |
| Maintenance Requirements | Low: Dry system, no additives or water pumps. | High: Constant monitoring of corrosion and seal replacements. |
| Corrosion Risk | Zero: Eliminates rust and mineral build-up. | High: Prone to scaling and internal corrosion. |
| Leak Impact | Minor: Non-critical for performance and easily repaired. | Critical: Causes thermal instability and safety hazards. |
| Weight & Logistics | Light: Easier rigging, installation, and lower shipping costs. | Heavy: Complex structural requirements for installation. |
Full Technical Specifications & Properties
BOCON Hot Air ETO Sterilizer System Details
| Feature Category | Technical Specification |
|---|---|
| STANDARDS & PHYSICAL ACCESS | |
| System Design Standards | EN1422, ISO11135-2014 |
| Door Configuration | 2 Doors; Pneumatic translation seal (Default); Optional: Rotary or Lifting |
| CONTROL SYSTEM & DATA INTEGRITY | |
| System Control Mode | Soft Manual & Full-Automatic |
| Data Recording | Interval mode & Warning triggered mode; Continuous curve recording |
| Data Storage Lifecycle | 10 Years (Secure HDD Storage) |
| Data Processing | PC Monitoring & Specialized Sterilization Data Software |
| SENSORS & PRECISION | |
| Temp Control Precision | Inside: ±3°C | Wall: ±5°C (Res: 0.1°C) |
| Pressure Management | Range: -100 to +150Kpa; Res: 0.1Kpa; Resistance: up to 80KPa (Standard) |
| Vacuum Performance | < 30 minutes (0 to -90KPa); Continuous/Stepping modes |
| Humidity Performance | 0-100%RH (non-condensing); Resolution: 0.1%RH |
| AUTOMATION & SAFETY CONNECTORS | |
| Automatic Detection | Cabinet water volume, Steam pressure, and Vacuum pump monitoring |
| Safety Alerts | Real-time display & warning for insufficient steam or pressure |
| System Protection | Nitrogen protection (Optional/Auto); Tail gas treatment (Auto-control) |
| Exhaust Control | In-door and Out-door air exhausting with automatic valve control |
| GAS RECIPES & PROCESS | |
| Supported Gas Recipes | Mixed Gas (CO2/EO mixtures: 70/30, 20/80, 10/90) and Pure EO (with N2) |
| Time Management | 0-99 hours; Resolution: 1 min |
Customized solutions for high-pressure chambers (>80KPa) and specific gas concentrations are available.