I-ETO Sterilization Real
I-ETO Sterilization Re-qualification Izidingo
1. Imvamisa yokufaneleka kabusha
•Isidingo esincane sonyaka
• Okungenani kanye ngonyaka, idatha yokuqinisekisa, idatha yokuqinisekisa kabusha elandelayo, kanye nedatha yokucubungula okujwayelekile kufanele kubuyekezwe.
• Ububanzi bokufaneleka kabusha kufanele bunqunywe futhi bubhalwe phansi
• Inqubo yokubuyekeza izobhalwa phansi
2. Ingabe Ukufaneleka Kabusha Kuyadingeka Uma Lungekho Ushintsho Lomkhiqizo Olwenzekayo?
•Yebo, ukufaneleka kabusha kusadingeka
3. Izinto zokufaneleka kabusha
•Ngokusekelwe kwimibandela ethile, izinketho ezilandelayo ziyatholakala:
- Ukufaneleka kabusha okugcwele
- Asikho isidingo sokuqinisekiswa kabusha ngokomzimba noma kwe-microbial
- Ukuqinisekiswa kokusebenza kwe-microbial encane
Izimfuneko zokufaneleka kabusha
a) Okulandelayo kufanele kubuyekezwe/kuhlolwe:
b) Noma ngabe ikhonaizinguquko ezibalulekile okuthinta ubunyumba bomkhiqizo kwenzeke ku-:
- Umklamo womkhiqizo
- Izinto zokwakha/zokupakisha
- I-PCD (Process Challenge Device)
- Abahlinzeki
- Izindawo/izinsiza zokukhiqiza
- Layisha ukucushwa noma izinqubo zokukhiqiza
c)Ukuhlaziywa kwethrendi ye-Bioburden kwemikhiqizo;
d) Nomaizifundo zokusabalalisa izinga lokushisa kanye nokusebenza kwesibulala-magciwane kukhombisa izinguquko ezibalulekile kusukela ekutholeni iziqu zokugcina;
e) Nomaukusatshalaliswa kwezinga lokushisa kanye namasheke okubuyisela kabusha emakamelweni e-preconditioning noma ezindaweni zokungenisa umoya abonise izinguquko ezibalulekile kusukela ekutholeni iziqu zokugcina;
f) Ukuthi i-ETO Sterilization umlando wokucubungula njengoba ukuqinisekiswa kokugcina kukhombisa ukukhiqiza kabusha;
g) Nomashintsha ukulawula kanye namarekhodi okulondoloza okuvimbela qinisekisa ukuthi akukho ukuguqulwa kwemishini ethinta inqubo;
h) Nomaazikho izinguquko kunqubo ye-ETO Sterilization okuthinta inzalo kwenzeke;
i) UmaUkucaciswa kwenqubo ye-ETO Sterilization kushintshile, ukufaneleka kabusha kufanele kubandakanye ukuqinisekiswa kweImikhawulo yensalela ye-EO ukuhambisana nezidingo zokulawula.
Amazinga Okulawula
- ISO 11135:2014 (Isiqephu 9.2: Imibandela yokufaneleka)
- I-FDA 21 CFR 211.113 (Izilawuli zenqubo yokuvala inzalo)
- I-EU GMP Isijobelelo 1:2022 (Isiqephu 8.5: Izidingo zokuqinisekisa)
I-ETO Sterilization Re-qualification Izidingo
Ukufaneleka kabusha kufanelehlanganisa ukubuyekezwa kokusebenza kwegumbi kanye nezinguquko zonyaka zobunjiniyela ukuze uqinisekise ukuthi imiphumela ye-IQ/OQ yasekuqaleni ihlala isebenza. Isibuyekezo kufanele sihlanganise:
- Amaphrofayli wokushisa/womswakama wezoni preconditioning (uma kusebenza);
- Amaphrofayela wangempela womswakama wegumbi elingenalutho;
- Amazinga okushisa endaweni yomoya (uma kusebenza);
- Ukwengeza, noma ngabe imininingwane yenqubo iyahlangatshezwa, ukuhlolwa kumele kucushwe yilokhu:
- Imikhuba emibi ekusebenzeni kwemishini,
- Ukuhluleka kokuzala;
- Ukunquma ukuthi ukuqinisekiswa kabusha kuyaqinisekiswa yini.
- Ochwepheshe bokuvala inzalo izochaza ububanzi bokudingekayoukuqinisekiswa kabusha ngokomzimba namagciwane ngokusekelwe emiphumeleni yokubuyekeza.
Ukucatshangelwa Okubalulekile Kokufaneleka Kabusha Kwenzalo
1. Ukusingatha Izinguquko Ezibalulekile
• Umaizinguquko ezinkulu zikhonjwa ngesikhathi sokufaneleka kabusha, kungase kudingeke kwenziwe kabusha i-IQ/OQ/PQ:
- I-IQ (Ukufaneleka Kokufaka): Qinisekisa ukufakwa kwemishini elungisiwe/ukufakwa kwesofthiwe.
- I-OQ (Ukufaneleka Kokusebenza): Hlola kabusha amapharamitha wokusebenza athintekile.
- I-PQ (Ukufaneleka Kokusebenza): Qinisekisa kabusha ukusebenza kokuvala inzalo ngaphansi kwezimo ezintsha.
2. Izidingo Ezengeziwe Zokukhishwa Kwepharamitha
Ngezinsiza ezisetshenziswayoukukhululwa kwe-parametric (kuncike kwidatha yenqubo esikhundleni sokuhlolwa kwe-BI):
a)Imvamisa yokuqinisekisa kabusha: Kumele kwenziweokungenani njalo ngonyaka;
b)Ububanzi bokuqinisekisa kabusha: Kumele ihlanganiseizifundo ze-microbiological ukuqinisekisa:
Izimo Ezidinga Ukufaneleka Okugcwele (IQ/OQ/PQ):
a)Ukugcinwa okuyinhloko kwe-sterilizer;
b)Izinguquko zesakhiwo, indawo, noma imvelo;
c)Ukuhluleka kobunyumba okungachazeki;
d)Izinguquko ku-EO supply/delivery, diluent, noma ukulungiselelwa kokulayisha kwegumbi.
Ukukhululwa ekuqeqesheni kabusha ngokomzimba noma kwe-Microbial
Lapho lezi zimo ezilandelayo zifinyelelwa:
- Azikho izinguquko emkhiqizweni, emaphaketheni, ezintweni zokusebenza/ezinsizeni, noma ezinqubweni;
- Ukusebenza kwegumbi nokubuyekezwa kobunjiniyela bafanelekile;
- Izinqubo zokuvala inzalo njalo asebenze ngokwethembeka phakathi nalesi sikhathi;
?Ukwahlulela kochwepheshe ingase inqume ukuthi ukufaneleka kabusha ngokomzimba noma kwe-microbialakudingekile kuze kube yisibuyekezo esilandelayo.
Ukufaneleka Kokusebenza Kwezinto Ezincane Ezilinganiselwe
•Kungase kudingeke ezimweni ezithile, njenge:
- Ukuqinisekisa ukuqhubeka kokufaneleka kwama-BI ngokuphathelene nomthwalo wezinto eziphilayo zomkhiqizo;
- Ukunikeza ubufakazi bokuthi chaizinguquko ezingahlosiwe kwenzeke kusukela ocwaningweni lwangaphambilini (kabusha) lweziqu ngemuva kwesikhawu sesikhathi esichaziwe.
•Izidingo ezijwayelekile (okungenani): - Eyodwaukugijima komjikelezo omfushane noma uhhafu womjikelezo;
- Kuhlanganisaizinga lokushisa kanye nomswakama izilinganiso wezinto ezifakwe inzalo.
Izibalo ezibalulekile
| Trigger | Scope | Ireferensi |
|---|---|---|
| Scheduled review | Documented periodic review of the validated state | ISO 11135:2014 §9.2 |
| No product change | Review still required; exemption from physical or microbial re-qualification only where justified | ISO 11135:2014 |
| Product or packaging change | Physical and microbial performance re-qualification | ISO 11135:2014 |
| Chamber or equipment change | Full IQ / OQ / PQ | ISO 11135:2014 |
| Limited-scale microbial PQ | Permitted under defined conditions when the change is bounded | ISO 11135:2014 |
| Regulatory basis | Sterilization process controls | FDA 21 CFR 211.113; EU GMP Annex 1:2022 §8 |
Imibuzo ejwayelekile ukubuzwa
How often must EO sterilization be re-qualified?
At least once per year. The annual review examines verification data, re-validation data and routine processing data, and the scope of the re-qualification must be determined and documented before the review starts.
Is re-qualification needed if no product changes occurred?
Yes. The annual review is independent of product changes — it exists to prove the process still performs as validated, using the year’s routine processing data, temperature distribution results, sterilizer operation records and preventive maintenance records.
Which items are covered by a re-qualification?
Chamber performance and annual engineering changes, preconditioning zone temperature and humidity profiles, empty-chamber humidity profiles, aeration zone temperatures, bioburden trend analysis, change-control and maintenance records, and the defined load configuration and PCD.
When is a full IQ/OQ/PQ repeated instead of a partial one?
After major maintenance of the sterilizer, structural, location or environmental changes, unexplained sterility failures, or changes in EO supply and delivery, diluent, or chamber load configuration. Other changes are handled by limited-scale microbial qualification or by review only.
Can physical or microbial re-qualification be waived?
Yes, when nothing changed in product, packaging, equipment, services or process, chamber performance and the engineering review are qualified, and routine sterilization has operated reliably through the period. Professional judgement then decides whether it can wait until the next review.