Ukuvalwa inzalo kwe-Ethylene Oxide (EO/ETO) kusalokhu kuyisisekelo semishini yezokwelapha ezwela ukushisa. Ukuqinisekisa ukuphepha, ukusebenza kahle, kanye nokuthobela imithetho, isibulala-magciwane kufanele sihlangabezane nezinkomba zokusebenza eziqinile. Ngezansi sihlukanisa ukucaciswa okubalulekile - kusukela ezintweni zokwakha kanye nezikhiye zeminyango ukuya ekulawuleni umjikelezo ozenzakalelayo kanye nokuqinisekisa.

1. Ukwakhiwa, Izinto & Ubukhulu

Ukubukeka & Isakhiwo

Ingaphandle lokubulala amagciwane kufanele libe bushelelezi, lifane ngombala, lingabi nama-burrs, imifantu, noma imihuzuko ebonakalayo. Zonke izinsimbi, amaswishi, namafindo kuzovikeleka futhi kusebenze ngaphandle kokuthikamezeka.

I-Chamber Dimensions

Amakamelo angama-rectangular achazwa ngobude, ububanzi, ukuphakama, nomthamo; amakamelo angama-cylindrical ngobubanzi, ukujula, kanye nevolumu. Isikhala esisebenzisekayo kufanele sigcine ibanga elincane lika-15 mm ukusuka ezindaweni ezishisayo kanye nama-30 mm ukusuka ezindaweni ezingashisiwe.

Ukuhambisana Kwezinto ezibonakalayo

Zonke izingxenye ezithinta i-EO, isitimu, amanzi, noma umoya ocindezelwe kufanele zimelane nokugqwala, zingasabeli nge-EO, futhi zingakhuthazi ukwenziwa kwe-polymerization. Irabha yemvelo kanye ne-latex akuvunyelwe. Izinto ezijwayelekile ezigunyaziwe: insimbi engagqwali (?8% nickel), i-ABS, i-PTFE, irabha ye-silicone.

2. Amapayipi, Isikhutha & Ukunikezela

Ukubhobhoza & Ukuxhumana

  • Amapayipi acwebileyo oketshezi kufanele angabi nazihibe futhi angavuzi.
  • Amalunga (ngaphandle kokuxhumeka kwesilinda se-EO) azoshiselwa.
  • Amapayipi aphethe i-gas sterilant noma isitimu adinga ukufakwa kwe-thermal.
  • Ishubhu eguquguqukayo ye-EO kufanele isebenzise insimbi engagqwali, i-PTFE-lined, noma irabha ye-nitrile.
  • Ukuhlonza okucacile kwesiqondisindlela esiphakathi nendawo nokugeleza kuyisibopho.

I-Vaporizer & Ukunikezwa kwe-EO

  • I-vaporizer ishisa i-EO ngaphambi kokungena ekamelweni; izindawo zokushisisa kufanele zihlanzeke.
  • Izinga lokushisa legesi ye-EO ngemva kokuhwamuka akufanele lidlule ku-70°C.
  • Ukushisisa okuqondile kwama-cartridge we-EO noma amasilinda akuvunyelwe.
  • Isihlungi (?50 ?m) sifakwe kulayini wokuhlinzeka nge-EO ku-vaporizer.
  • Umthamo osuselwe esisindweni: ukunemba kwesikali phakathi kuka-±5% wenkokhiso ye-EO.
3. Izinhlelo Zomnyango & I-Safety Interlocks

Ukuvalwa Komnyango & Ukufinyelela

Iminyango kufanele ibe nezimpawu ezishintshwayo, ezihlanzeka ngaphandle kokukhipha umnyango. Ngaphambi kokuqala komjikelezo, umnyango wokulayisha ungavulwa. Uma umjikelezo uqala, umnyango uhlala uvaliwe kuze kube “umjikelezo uphelile” - ukuvula kudinga ukhiye ozinikele, ikhodi, noma ithuluzi.

Okuzenzakalelayo & Iminyango Yezandla

Iminyango ezenzakalelayo ihlala ikhiyiwe futhi ivaliwe uma amandla ehluleka. Amandla okusebenza kwesicabha akumele adlule ku-150 N. Izikhali zokukhiya ama-radial zidinga unogada okhiphekayo ukuze zihlolwe.

Izibulala-magciwane ezinezicabha ezimbili

Ukuvuleka ngesikhathi esisodwa kokulayisha nokuthulula iminyango kuyavinjelwa ngaphandle kwesondlo. Izinkomba nhlangothi zombili (kanye negumbi lokulawula kude) zibonisa ukuthi yimuphi umnyango ongavulwa.

4. Izinsimbi & Ukulawula Okuzenzakalelayo

Amathuluzi ayimpoqo afaka:

  • Inkomba yengcindezi yekamelo
  • Inkomba yezinga lokushisa lekamelo
  • Inkomba yesigaba sokuvala inzalo
  • Inkomba ephelele yomjikelezo
  • Inkomba yesimo somnyango
  • Isikhombi sokushisa kwejakhethi
  • I-EO inlet lokushisa (uma umjovo >30s)
  • Izinga lokushisa & amarekhodi okucindezela
  • Imonitha yengcindezi yokuvuza

Izinga lokushisa & Ukunemba Kwengcindezi

Izinga lokushisa: ububanzi 0–100°C, ukunemba ±1%, ukulungiswa 1°C, iphutha lenkomba <±1°C.
Ingcindezi: ububanzi obungu-0 kPa ukuya kungcindezi yokusebenza ephezulu, ukunemba okungu-±1.6%, iphutha lokucindezela kokuvala inzalo <±5 kPa. Imonitha yokuhlola ukuvuza: 20 kPa span, ±1% ukunemba, 0.1 kPa ukulungiswa.

Iyaqopha & Ubuqotho Bedatha

Amarekhodi ahlala njalo ayadingeka (alikho iphepha elishisayo elifiphalayo). Isikhawu samasampula ?imizuzwana emi-2. Amarekhodi edijithali kufanele abe nobufakazi bokuphazamiseka. Amashadi e-analog: isivinini ?1 mm/min, ububanzi ?100 mm. Amanani aqala/aphela omjikelezo wokuphrinta amarekhodi edijithali anezaziso zokungabekezelelani.

Isilawuli esizenzakalelayo

Izibulala-magciwane ze-Class A zivumela ukuhlelwa komsebenzisi ngokhiye/ikhodi; Ikilasi B lihlelwe ngokwefekthri kuphela. Ukuvikelwa kwezinga lokushisa elingaphezu kwamandla ngaphandle kwesistimu yokulawula. Isoftware kufanele iqinisekiswe. Amaphutha abangela i-alamu ebonakalayo/ezwakalayo kanye nesimo esiphephile.

5. Izigaba zomjikelezo wokuvala inzalo (9 Izigaba)
  1. I-Chamber pre-heating
  2. Isigaba se-vacuum
  3. Ukuhlolwa kokuvuza okuzenzakalelayo
  4. I-conditioning (ukushisa & thambisa)
  5. Umjovo we-EO
  6. Ukuvezwa kwenzalo
  7. Uyayiqeda
  8. Ukungenisa umoya / ukuwasha
  9. Ukungena emoyeni & ukuphela komjikelezo

Izidingo ezibalulekile: Ukuhlolwa kokuvuza kufanele kudlule (ushintsho lokucindezela ?0.3 kPa/min). Isimo Sekilasi B kufanele silondoloze umswakama ohlobene >40%. Isikhathi sokuvala inzalo siqala kuphela lapho izinga lokushisa, ingcindezi, nokugxilisa ingqondo kufinyelela ezindaweni ezimisiwe. Ngemuva komjikelezo, uma umnyango ungavulwa ngokushesha, ukungena kwe-aeration okwengeziwe noma ukushaywa okuqhubekayo kufanele kuqinisekise amazinga e-EO aphephile.

6. Ukusebenza, Ukuphepha kukagesi & I-EMC
  • Izinga lokushisa elingaphezulu kwegumbi ? izinga lokushisa +5°C.
  • Ukufana kwezinga lokushisa kwegumbi elingenalutho: ±3°C we-setpoint.
  • Isistimu ye-vacuum kufanele ifinyelele -75 kPa noma ngaphansi.
  • Izinga lokuvuza: ?0.1 kPa/min kumavolumu ?1 m³.
  • Izinga lomsindo ?65 dB(A).
  • Ukufakwa kwe-asbestos-free; indawo engaphezulu kuka-55°C kumele ivalwe.
  • Ukungasebenzi kwenkomba yebhayoloji kuyadingeka kuzibulala magciwane Zekilasi B.
  • Ukuphepha kukagesi nge-GB4793.1 / IEC 61010-2-040.
  • Ukuhambisana kwe-electromagnetic nge-GB/T18268.1.
  • Ukuqinisekiswa kwe-sterilant ngaphandle kokulawulwa komjovo (ingcindezi, isisindo, noma ivolumu).
  • Izinyathelo zokuxilongwa kwephutha zokunakekela.

Ufuna isibulali magciwane se-ETO esithobelanayo?

Ama-sterilizer ethu e-ethylene oxide ahlangabezana nokucaciswa kokusebenza okuqinile. Xhumana nathi namuhla ukuze uthole ukubonisana nobuchwepheshe noma ukucaphuna.

Xoxa ku-WhatsApp

Umshwana wokuzihlangula: Lesi sihloko sifinyeza izinkomba zokusebenza ngezinjongo zemfundo. Ngaso sonke isikhathi bheka imibhalo yomkhiqizi kanye nemithethonqubo esebenzayo yezidingo ezithile.

Skrolela phezulu