3.4.1 Umshini we-eto sterilizer Ububanzi be-Temperature Setting
Ibanga lokusetha izinga lokushisa le-eto sterilizer lithi: 30 kuya ku-70°C, nephutha lokulawula elingu-?2°C.
3.4.2 Izinga lokushisa Ukufana komshini we-ethylene oxide
Ukufana kwezinga lokushisa kwe-eto sterilizer kufanele kuhlangabezane nezidingo ezilandelayo:
Uma ivolumu yegumbi lokuvala inzalo ingu-?5m³, okungenani amaphoyinti okulinganisa ayi-10 kufanele abekwe.
Lapho ivolumu >5m³, ku-1m³ ngayinye eyengeziwe, iphoyinti lokulinganisa elilodwa kufanele lengezwe.
Uma ivolumu ingu-?10m³, okungenani amaphoyinti okulinganisa angu-20 kufanele asethwe ngaphakathi kwegumbi, nezinga lokushisa lokuhlola elingu-?40°C, futhi umehluko wezinga lokushisa phakathi kwephoyinti ngalinye lokulinganisa akumele udlule ku-±3°C.
3.4.3 Izinga Lokuvuza Kwegumbi Lokuvala Inzalo
3.4.3.1
Ukuhlolwa kufanele kwenziwe ngaphambi kwesigaba somjovo we-ejenti ebulalayo kanye nesigaba sokumswakama (uma kusetshenziswa izinga lokushisa elifakiwe kanye nedivayisi yokubonisa ukucindezela). Ingcindezi kufanele imiswe kusengaphambili ngumkhiqizi, futhi ingcindezi yokuhlola kufanele ibe ngaphansi kuka -20 kPa futhi okungenani ibe ngu-20 kPa ibe ngaphansi kunokucindezela okusethiwe kokuvala inzalo.
3.4.3.2
Uma ingcindezi phakathi kwanoma yisiphi isigaba somjikelezo wokuvala inzalo ingaphansi komfutho womoya, ngemva kokufinyelela ukucindezela osethiwe, wonke amavalvu axhunywe egumbini lokuvala inzalo kufanele avalwe, futhi iphampu yevacuum kufanele iyeke ukusebenza. Ukucindezela ekamelweni lokuvala inzalo kufanele kuqashelwe okungenani imizuzu emi-5, lapho ukuphakama kwengcindezi akufanele kudlule ku-0.3 kPa/min.
3.4.3.3
Uma ukucindezela kokusebenza kwegumbi le-eo kungaphezu kuka-5 kPa, ukuhlolwa kokuvuza kwengcindezi enhle kufanele kwenziwe, okufanele kwenziwe ngaphambi kwesiteji se-vacuum. Ngokucindezela ngegesi ye-inert (njengomoya) ukuya ku-10% wengcindezi yokusebenza esethiwe, ukucindezela egumbini lokuvala inzalo kufanele kuqashelwe okungenani imizuzu emi-5, lapho ukwehla kwengcindezi akufanele kudlule ku-0.3 kPa/min.
3.4.4 Isilinganiso seVacuum yomshini we-eto
Ngaphansi kwezimo zokungalayishi, i-eo chamber kufanele ifinyelele -15 kPa ukusuka kumfutho womkhathi ngemizuzu engu-?6; finyelela -50 kPa ngemizuzu engu-30; futhi ifinyelele -75 kPa ngemizuzu engu-40.
3.4.5 I-Humidification Umphumela womshini we-eto sterilizer
Ngaphansi kwezimo zokungalayishi, lapho ukuswakama okuhlobene kungu-?40% RH, umshini wokukhipha inzalo i-eo kufanele ube nekhono lokukhulisa umswakama ohlobene ufike okungenani ku-75% noma ngaphezulu ngohlelo lokuswakama.
3.4.6 UHlelo Lokubeka Ingcindezi Yegumbi Lokusterilisa
Ibanga lokusetha ingcindezi legumbi lokuvala inzalo lingu--0.08 MPa ukuya ku-+0.08 MPa.
3.4.7 Izinga Lomsindo Elikhulu lomshini we-eto sterilizer
Izinga eliphezulu lomsindo (A-weighted) elikhiqizwe i-eto sterilization chamber kufanele libe ngu-?65 dB.
3.4.8 Ukubukeka kanye Nesakhiwo se-ethylene oxide sterilizer
3.4.8.1
Umumo kufanele uqonde, ube nendawo ebushelelezi futhi ehlanzekile, umbala ofanayo, futhi ungabi nama-burrs, imiphetho ebukhali, nemifantu. Akufanele kube khona imihuzuko esobala noma amaphutha njengamaqhubu.
3.4.8.2
Wonke umbhalo, imidwebo, izimpawu, njll., endaweni kufanele iphrintwe ngokucacile, ibe nezimpawu ezicacile, ezinembile, nezihlala isikhathi eside.
3.4.8.3
Izinto zokubopha kufanele zifakwe ngokuvikelekile, futhi wonke amaswishi okulawula namafindo okulungisa (izinkinobho) kufanele asebenze ngendlela eguquguqukayo futhi enokwethenjelwa, ngaphandle kokuphazamiseka.