Gesuiwerde watervereistes vir ETO-sterilisasie

Process Water For ETO Sterilization Equipment

Die bevogtigingswater wat in die ETO-sterilisasietoerusting gebruik word, moet vry wees van kontaminante, moet nie die ETO-sterilisasieproses verswak nie, en moet nie die sterilisator, gesteriliseerde produkte of die vrag beskadig nie. Bevogtiging moet gedoen word deur stoom in te spuit, nie deur direk vernevelde water in te spuit nie.

Note: It is recommended to use purified water that meets the requirements of the Chinese Pharmacopoeia for humidification. If purified water is not used, the steam supply or steam condensate should be validated to ensure that the humidification steam meets the requirements of this section. The steam quality testing methods are outlined in YY/T 1612 and the Chinese Pharmacopoeia.

Steam Contaminant Limits

ContaminantCondensateSteam Supply
ResidueLess than 10mg/kgLess than 100mg/L
Heavy MetalsNot exceeding the limits for purified water in the Chinese PharmacopoeiaNot exceeding the limits for purified water in the Chinese Pharmacopoeia
ChloridesLess than 1mg/kgLess than 10mg/L
ConductivityLess than 15?S/cmLess than 50?S/cm
pH5–8.55–8.5
AppearanceColorlessColorless

Note 1: If the contaminant levels in the feedwater or steam exceed the above limits, the sterilizer’s service life will be significantly reduced.

Process Air For Sterilization

The air used in the sterilizer should be dried and filtered to ensure the service life and filtering performance of the ETO Sterilization sterilizer’s terminal filter.