Process Water For ETO Sterilization Equipment

El agua de humidificación utilizada en el equipo de esterilización ETO debe estar libre de contaminantes, no debe debilitar el proceso de esterilización ETO y no debe dañar el esterilizador, los productos esterilizados ni la carga. La humidificación se debe realizar inyectando vapor, no inyectando directamente agua atomizada.

Note: It is recommended to use purified water that meets the requirements of the Chinese Pharmacopoeia for humidification. If purified water is not used, the steam supply or steam condensate should be validated to ensure that the humidification steam meets the requirements of this section. The steam quality testing methods are outlined in YY/T 1612 and the Chinese Pharmacopoeia.

Steam Contaminant Limits

ContaminantCondensateSteam Supply
ResidueLess than 10mg/kgLess than 100mg/L
Heavy MetalsNot exceeding the limits for purified water in the Chinese PharmacopoeiaNot exceeding the limits for purified water in the Chinese Pharmacopoeia
ChloridesLess than 1mg/kgLess than 10mg/L
ConductivityLess than 15?S/cmLess than 50?S/cm
pH5–8.55–8.5
AppearanceColorlessColorless

Note 1: If the contaminant levels in the feedwater or steam exceed the above limits, the sterilizer’s service life will be significantly reduced.

Process Air For Sterilization

The air used in the sterilizer should be dried and filtered to ensure the service life and filtering performance of the ETO Sterilization sterilizer’s terminal filter.