we need to prepare berfore the eo sterilization

  1. Appearance Inspection: Check for any damage on the surfaces of the main machine, control rack, and electrical control cabinet.
  2. Power Supply Check: Ensure the power supply meets the requirements and check the rotation direction of the water pump and vacuum pump.
  3. Куатым чӱктыза да эскерыза, приборым ончыктымо значений-влак тыглай улыт мо; уке гын, куатым кораҥдыза, угыч тӱҥалме деч ончыч тергыза да олмыктыза.
  4. Pipeline Inspection: Check for leaks in the water supply, vacuum, and medication delivery pipelines, and ensure all valves are in standby mode; adjust if necessary.
  5. Set Parameters: For example, set the dosage amount. Place the medication bottle on an electronic scale (provided by the user), confirm the bottle weight, start the vacuum pump to reach the set pressure value, then open the medication bottle to deliver the medication. Generally, the dosage during eo eo Sterilization is 400–800 mg/L (referring to the pure ethylene oxide input; if it is a mixed gas containing ethylene oxide, ensure that the amount of ethylene oxide delivered meets the dosage requirements). When the delivered weight reaches the standard, the current pressure value corresponds to the dosage amount.
  6. Each time before eo Sterilization, verify that the parameter settings meet the needs for the items being sterilized; otherwise, reset the eo Sterilization parameters.

IF YOU WANT TO KNOW MORE About Ethylene Oxide Sterilization