EO sterilization equipment and process characteristic patterns and effect analysis

Start cycle phase(Select parameters/sterilization recipe)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Choosing the wrong recipeLack of care in choosing the formulationInvalid cycle. Affects product sterility (re-sterilization)1. Different recipe names6Нэвтрэлт 5103001. Work instructions 2. Careful inspection before the cycle starts 3. Operator training2Нэвтрэлт 54401. Work instructions 2. Careful inspection before starting the cycle 3. Operator traininguser
Loss of product/packaging functionality and integrity due to unvalidated cycle parameters1. Different recipe names64102401. Work instructions 2. Careful inspection before the cycle starts 3. Operator training244321. Work instructions 2. Careful inspection before starting the cycle 3. Operator traininguser
Warm-up phase(Cabinet heating)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensationНэвтрэлт 53 дайсан2301. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Automatic heating time compensation63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)
1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm
6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
2The temperature inside the cabinet is too highTemperature sensor failureAffects product sterility (re-sterilization)
1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm
642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
Insulation stage(Maintain cabinet temperature)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarmНэвтрэлт 53 дайсан2301. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Heating timeout alarm 2. Circulating water pump overload alarm6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
2The temperature inside the cabinet is too highTemperature sensor failureAffects product sterility (re-sterilization)1. Heating timeout alarm 2. Circulating water pump overload alarm642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
3 дайсанLack of timeProgramming ErrorsRepeat sterilization1. Automatic cycle inspection24216none
Pre-vacuum stage(Removing air from the chamber allows the process to proceed below the EO flammability limit)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Not enough air removedLow vacuum pump efficiencyThe cycle is locked in one phase until it is restored and restarted after a standby period1. Pre-vacuum time over limit alarm 2. Vacuum pump overload alarm 3. Automatically detect water flow and shut down the waterless pumpНэвтрэлт 53 дайсан2301. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Wrong valveThe cycle is locked in one phase until it is restored and restarted after a standby period1. Pre-vacuum time over limit alarm 2. Vacuum pump overload alarm63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Pressure sensor failureFlammable Zone Hazard1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit63 дайсан2361. Preventive maintenance 2. Calibration23 дайсан2121. Preventive maintenance 2. Calibrationuser
2Too much vacuumPressure sensor failureProduct and/or packaging damage (classification)1. Redundant pressure sensor 2. Pressure difference alarm is issued when the pressure difference is greater than 1kpa.642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
Leak detection stage(Check the tightness of the box)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Cabinet leakageGasket leakageNo impact1. Leak test failure alarm 2. Process ends10Нэвтрэлт 521001. Preventive maintenance3 дайсанНэвтрэлт 52301. Preventive maintenanceuser
Valve leakageNo impact1. Leak test failure alarm63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Circulation system air duct leakageNo impact1. Leak test failure alarm43 дайсан2241. Preventive maintenance23 дайсан2121. Preventive maintenance 2. Calibrationuser
Nitrogen replacement stage(Fill the chamber with an inert gas to replace oxygen)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Not enough nitrogen entering the chamberInsufficient nitrogenThe cycle is locked in one phase until it is restored and restarted after a standby periodEnd of process3 дайсан3 дайсан218N/A
Faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby periodN/A63 дайсан3 дайсан541. Preventive maintenance23 дайсан3 дайсан181. Preventive maintenanceuser
Pressure sensor failureFlammable Zone Hazard1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
2Too much nitrogen in the cabinet doorFaulty valveEO concentration is lower than the validated value because during EO injection, if controlled by pressure (resterilization), additional N2 is injected1. Indoor high voltage alarm 2. End of process6Нэвтрэлт 541201. Preventive maintenance2Нэвтрэлт 54401. Preventive maintenanceuser
Pressure transmitter failureToo much EO gas injected. High EO residue. Extended degassing time me)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
3 дайсанThe cabinet temperature is higher than the upper limit tImpact on product function (sorting)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
4The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Automatic heating time compensation63 дайсан3 дайсан541. Preventive maintenance23 дайсан3 дайсан181. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
Low carburetor temperatureAffects product sterility (re-sterilization)1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation642481. Preventive maintenance242161. Preventive maintenanceuser
Humidification stage(Increasing the humidity of the load by injecting steam)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Insufficient steamSteam generator failureAffects product sterility (re-sterilization)1. Alarm when steam pressure is lower than the limit6Нэвтрэлт 52601. Preventive maintenance2Нэвтрэлт 52201. Preventive maintenanceuser
Valve failureThe cycle is locked in one phase until it is restored and restarted after a standby period1.Humidification frequency limit alarm6Нэвтрэлт 52601. Preventive maintenance2Нэвтрэлт 52201. Preventive maintenanceuser
Pressure transmitter/humidity sensor failureAffects product sterility (re-sterilization)1. Redundant pressure/humidity sensor 2. Alarm when pressure/humidity difference exceeds limit6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
Filter cloggedThe cycle is locked in one phase until it is restored and restarted after a standby period1.Humidification frequency limit alarm6Нэвтрэлт 52601. Preventive maintenance2Нэвтрэлт 52201. Preventive maintenanceuser
2Too much steam injectionFaulty valveExcessive steam injection, wet cartons (sorting)1. Indoor high voltage alarm 2. End of process6Нэвтрэлт 541201. Preventive maintenance2Нэвтрэлт 54401. Preventive maintenanceuser
Pressure transmitter/humidity sensor failureExcessive steam injection, wet cartons (sorting)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
3 дайсанTemperature is too highTemperature sensor failureImpact on product functions (ranking)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
4The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked in one phase until it is restored and restarted after a standby period1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation63 дайсан2361. Preventive maintenance242161. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked in one phase until it is restored and restarted after a standby period1. Heating timeout alarm 2. Automatic heating time compensation63 дайсан3 дайсан541. Preventive maintenance23 дайсан3 дайсан181. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Heating timeout alarm 2. Circulating water pump overload alarm6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
Humidification dwell phase(Balanced humidity)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Lack of timetime outAffects product sterility (re-sterilization)1. Automatic cycle inspection2Нэвтрэлт 52201. Preventive maintenance1-1Нэвтрэлт 52101. Preventive maintenanceuser
Nitrogen cushion stage(Fill the chamber with inert gas to dilute the oxygen)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Not enough nitrogen entering the cabinetLack of nitrogenThe cycle is locked in one phase until it is restored and restarted after a standby period1. Check N2 pressure before starting the cycle 2. Alarm when N2 pressure is lower than the limit3 дайсан3 дайсан2180
Valve failureThe cycle is locked in one phase until it is restored and restarted after a standby periodnone63 дайсан3 дайсан541. Preventive maintenance23 дайсан3 дайсан18user1. Preventive maintenance
Pressure sensor failureFlammable Zone Hazard1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit63 дайсан2361. Preventive maintenance23 дайсан212user1. Preventive maintenance
2Too much nitrogen injected into the cabinetFaulty valveEO concentration is lower than the validated value because during EO injection, if controlled by pressure (resterilization), additional N2 is injected1. Indoor high voltage alarm 2. End of process6Нэвтрэлт 541201. Preventive maintenance2Нэвтрэлт 5440user1. Preventive maintenance
Pressure transmitter failureToo much EO gas injected. High EO residue. (Extend degassing time)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 5220user1. Preventive maintenance 2. Calibration
3 дайсанThe temperature inside the cabinet exceeds the upper limitTemperature sensor failureImpact on product function (sorting)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit642481. Preventive maintenance 2. Calibration24216user1. Preventive maintenance 2. Calibration
4The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation63 дайсан2361. Preventive maintenance23 дайсан212user1. Preventive maintenance
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation63 дайсан3 дайсан541. Preventive maintenance23 дайсан3 дайсан18user1. Preventive maintenance
Temperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 5220user1. Preventive maintenance 2. Calibration
Low carburetor temperatureAffects product sterility (re-sterilization)1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation642481. Preventive maintenance 2. Calibration24216user1. Preventive maintenance 2. Calibration
EO injection stage(Fill the cabinet with EO gas)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Not enough ethylene oxide is entering the cabinetMissing EOThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1.EO cylinder hole alarm 2.EO injection timeout alarm10Нэвтрэлт 521001. Operator training 2. EO weight check before cycle start2Нэвтрэлт 52201. Operator training 2. EO weight check before cycle startuser
Faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby period1.EO injection timeout alarm6Нэвтрэлт 52601. Preventive maintenance6Нэвтрэлт 52601. Preventive maintenanceuser
EO pipeline leakageBatches that exceed the qualification period may be sterilized againnone6Нэвтрэлт 53 дайсан901. Preventive maintenance 2. LEL EO detector alarm2Нэвтрэлт 52201. Preventive maintenance 2. LEL EO detector alarmuser
Pressure sensor failureEO concentration is lower than the validation value (re-sterilization)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
2Liquid EO enters the cabinetLow carburetor temperatureThe cycle is locked in one phase until it is restored and restarted after a standby period1. The carburetor temperature is lower than the limit alarm642481. Preventive maintenance242161. Preventive maintenanceuser
Not enough water in the carburetor due to a faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby period1. The carburetor water level is lower than the limit alarm642481. Preventive maintenance242161. Preventive maintenanceuser
3 дайсанToo much ethylene oxide input in the cabinetFaulty valveLonger quarantine, requiring additional testing1. Indoor high voltage alarm63 дайсан3 дайсан541. Preventive maintenance 2. EO residual inspection23 дайсан3 дайсан181. Preventive maintenance 2. EO residual inspectionuser
Pressure sensor failureLonger quarantine, requiring additional testing1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit63 дайсан2361. Preventive maintenance 2. Calibration23 дайсан2121. Preventive maintenance 2. Calibrationuser
4The temperature inside the cabinet exceeds the upper limitTemperature sensor failureImpact on product function (sorting)1. Redundant temperature sensor 2. Chamber high temperature alarm 3. Temperature difference exceeds limit alarm642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
Нэвтрэлт 5The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Automatic heating time compensation63 дайсан3 дайсан541. Preventive maintenance23 дайсан3 дайсан181. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
Low carburetor temperatureAffects product sterility (re-sterilization)1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation642481. Preventive maintenance242161. Preventive maintenanceuser
Nitrogen cap stage(Add N2 to cover EO)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Not enough nitrogen entering the cabinetLack of nitrogenThe cycle is locked in one phase until it is restored and restarted after a standby period1. Check N2 pressure before starting the cycle 2. Alarm when N2 pressure is lower than the limit3 дайсан3 дайсан218N/A
Valve failureThe cycle is locked in one phase until it is restored and restarted after a standby periodnone63 дайсан3 дайсан541. Preventive maintenance23 дайсан3 дайсан181. Preventive maintenanceuser
Pressure sensor failureFlammable Zone Hazard1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
2Too much nitrogen injected into the cabinetFaulty valveThe cabin is under positive pressure. This cycle will end.1. Indoor high voltage alarm 2. End of process6Нэвтрэлт 541201. Preventive maintenance2Нэвтрэлт 54401. Preventive maintenanceuser
3 дайсанThe temperature inside the cabinet exceeds the upper limitTemperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Chamber high temperature alarm 3. Temperature difference exceeds limit alarm642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
4The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Automatic heating time compensation63 дайсан3 дайсан541. Preventive maintenance23 дайсан3 дайсан181. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
Low carburetor temperatureAffects product sterility (re-sterilization)1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation642481. Preventive maintenance242161. Preventive maintenanceuser
EO exposure stage(Disinfect the product)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Lack of timeProgramming ErrorsRepeat sterilization1. Automatic cycle inspection2Нэвтрэлт 52201. Preventive maintenance1-1Нэвтрэлт 52101. Preventive maintenanceuser
2The box temperature is lower than the lower limit tHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.Heating timeout 2. Circulating water pump overload alarm 3. Automatic heating time compensation63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked in one phase until it is resumed and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Automatic heating time compensation63 дайсан3 дайсан541. Preventive maintenance23 дайсан3 дайсан181. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm6Нэвтрэлт 52601. Preventive maintenance 2. Calibration2Нэвтрэлт 52201. Preventive maintenance 2. Calibrationuser
3 дайсанThe temperature inside the cabinet exceeds the upper limitTemperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Chamber high temperature alarm 3. Temperature alarm642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
4Not getting a sufficiently homogeneous EO mixtureCirculation fan failureAffects product sterility (re-sterilization)1. Low speed alarm6Нэвтрэлт 53 дайсан901. Preventive maintenance2Нэвтрэлт 53 дайсан301. Preventive maintenanceuser
Product positioning very close to the cavity wallAffects product sterility (re-sterilization)1.Guardrail is used to prevent the product from contacting the cavity wall2Нэвтрэлт 53 дайсан300N/A
EO Removal Phase(Create vacuum/remove sterilization gas)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Not purging enough gasInefficient vacuum pumpThe cycle is locked in one phase until it is restored and restarted after a standby period1.EO removal timeout alarm 2.Vacuum pump overload alarm 3.Automatically detect water flow and shut down the waterless pump63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby period1.EO removal timeout alarm 2.Vacuum pump overload alarm63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Pressure sensor failureLonger isolation time and additional testing required due to less EO gas being removed1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit63 дайсан2361. Preventive maintenance 2. Calibration23 дайсан2121. Preventive maintenance 2. Calibrationuser
2Exhaust too much gasPressure sensor failureProduct and/or packaging damage (classification)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
Nitrogen purge stage(Ensure that the indoor environment is not flammable)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Not enough nitrogen entering the chamberLack of nitrogenThe cycle is locked in one phase until it is restored and restarted after a standby period1. Check N2 pressure before starting the cycle 2. Alarm when N2 pressure is lower than the limit3 дайсан3 дайсан218N/A
Faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby periodnone63 дайсан3 дайсан541. Preventive maintenance23 дайсан3 дайсан181. Preventive maintenanceuser
Pressure sensor failureFlammable Zone Hazard1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Longer quarantines require additional testing1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Air cleaning stage(Create vacuum/remove residual sterilization gas, eliminate residual EO)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Not purging enough gasInefficient vacuum pumpThe cycle is locked in one phase until it is restored and restarted after a standby period1.EO removal timeout alarm 2.Vacuum pump overload alarm 3.Automatically detect water flow and shut down the waterless pump63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby period1.EO removal timeout alarm 2.Vacuum pump overload alarm63 дайсан2361. Preventive maintenance23 дайсан2121. Preventive maintenanceuser
Pressure sensor failureLonger isolation time and additional testing required due to less EO gas being removed1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit63 дайсан2361. Preventive maintenance 2. Calibration23 дайсан2121. Preventive maintenance 2. Calibrationuser
2Exhaust too much gasPressure sensor failureProduct and/or packaging damage (classification)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
3 дайсанInsufficient air in the cabinFaulty valveThe cycle is locked in one phase until it is restored and restarted after a standby period1. Gas injection time over limit alarm643 дайсан721. Preventive maintenance243 дайсан241. Preventive maintenanceuser
Filter cloggedThe cycle is locked in one phase until it is restored and restarted after a standby period1. Gas injection time over limit alarm643 дайсан721. Preventive maintenance243 дайсан241. Preventive maintenanceuser
Pressure sensor failureLonger quarantines require additional testing1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit63 дайсан2361. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
Removal stage(Remove residue from the product)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1-1Aeration time is less than the specified timeProgramming ErrorsRepeat sterilization1. Automatic cycle inspection23 дайсан2121. Restart your computer regularly1-13 дайсан261. Restart your computer regularlyuser