Introduction
Ethylene Oxide (EO) sterilization remains a preferred method for medical devices incompatible with radiation or moist heat. However, it’s a top source ofquality management system deficiencies during FDA/ISO audits. Proactively addressing these non-conformities is essential forsterile medical device registration and compliance.

I. Sterilization Validation Failures
Critical gaps in proving process effectiveness:
- ?Undefined processes: Missing sterilization parameters in design documentation.
- ?SAL omissions: Sterility Assurance Level (SAL) not specified in reports.
- ?Unapproved/incomplete reports: Lack evaluative conclusions or raw test data.
- ?Software validation gaps: No data for sterilizer software systems.
- ?Revalidation issues: Triggers/cycles undefined; changes made without revalidation.
- ?Parameter mismatches: Conflicts between protocols, records, or physical equipment IDs.
- ?Aeration flaws: Unclear conditions/duration in documentation.
?Fix: Align with ISO 11135. Validate ALL critical parameters (load configs, BI placement, gas dosage).
II. Warehouse & Storage Risks
EO safety and material control violations:
- ?Poor facility controls: Weeds, stagnant water, or debris in production zones.
- ?Inadequate aeration: Missing/undersized rooms; poor ventilation or exhaust.
- ??Unsafe EO storage: Non-dedicated zones; cylinders stored haphazardly without:
- Temperature/humidity monitoring
- Fire controls or leak detection
- ?Material management gaps: Unlabeled sterile items; no environmental monitors.
III. Sterilization Process Control Errors
Deviations during execution:
- ?Poor traceability: Ambiguous batch numbering; missing device IDs/logs.
- ??Missing critical parameters in logs:
- Gas injection time, EO dosage, vacuum cycles
- BI lot numbers, temperature/humidity data
- ?Uncontrolled deviations: Parameters not matching validated specs or work instructions.
- ?Missing safety systems: No gas leak detectors or load diagrams in SOPs.
IV. Residual Control Oversights
Patient safety risks from EO residues:
- ?Missing validation: No data supporting residue removal methods.
- ?Incomplete records: Raw test data or equipment IDs absent.
- ??Non-compliant limits: Unsubstantiated EO residue thresholds.
V. Documentation & Personnel Gaps
Systemic compliance weaknesses:
- ?Uncontrolled documents: Obsolete SOPs; unnumbered forms; unsigned records.
- ???Equipment neglect: Using faulty sterilizers; unlabeled piping; expired calibrations.
- ???Untrained staff: No competency assessments for sterilization operators.
VI. Supply Chain & Sales Issues
Traceability breakdowns:
- ?Incomplete sales records: Missing batch IDs, expiration dates, or client details.
- ?Expired contracts: EO service agreements not renewed.
- ?Sample deviations: Retained sample quantities mismatched with procedures.