Imigomo nezincazelo ze-eto gas inzalo
For the purposes of this document, the following terms and definitions apply in ethylene oxide sterilization 3.1 Aeration part of […]
Imigomo nezincazelo ze-eto gas inzalo Funda kabanzi
For the purposes of this document, the following terms and definitions apply in ethylene oxide sterilization 3.1 Aeration part of […]
Imigomo nezincazelo ze-eto gas inzalo Funda kabanzi
Ukuhlola Ukubukeka: Hlola noma yimuphi umonakalo endaweni yomshini omkhulu, indawo yokulawula, kanye nekhabhinethi yokulawula kagesi.
Ukuhlola Ukunikezwa Amandla: Qinisekisa ukuthi ugesi uhlangabezana nezidingo futhi uhlole indlela yokuzungezisa iphampu yamanzi kanye nephampu yevacuum.
Vula amandla bese ubheka ukuthi amanani ensimbi ajwayelekile yini; uma kungenjalo, nqamula amandla, hlola futhi ulungise ngaphambi kokuqala kabusha.
Ukuhlolwa Kwepayipi: Hlola ukuvuza kwamanzi, i-vacuum, namapayipi okulethwa kwemithi, futhi uqinisekise ukuthi wonke ama-valve akumodi yokulinda; lungisa uma kunesidingo.
Umsebenzi Wokulungiselela Ngaphambi Kokwenza Inzalo Funda kabanzi
2.1.1 Umumo jikelele kufanele unembe, ube nezindawo ezibushelelezi, ezihlanzekile nombala ofanayo. Akufanele kube nama-burrs, imiphetho ebukhali, imifantu, imihuzuko ebonakalayo, noma ukungalingani.
2.1.2 Wonke umbhalo wangaphandle, ihluzo, nezimpawu kumele ziphrintwe ngokucacile, zibe nezimpawu ezinembile neziqinile.
2.1.3 Izinsimbi kufanele zifakwe ngokuphephile. Wonke amaswishi okulawula namafindo okulungisa (noma izinkinobho) kufanele asebenze kahle nangokwethembeka ngaphandle kwezithiyo.
Isicelo Somkhiqizi we-Eto Gas Sterilizer Funda kabanzi
Ngaphakathi kohlu oluthile lokuhlushwa kwegesi, ukugcwala okuphezulu, umphumela ongcono we-EO Gas Sterilization; ngaphezu kuka-1000 mg/L, ukuthuthukiswa komphumela wokuvala inzalo akubalulekile.
Izinto Ezithinta Inqubo Yokuvala Inzalo Yegesi ye-EO Funda kabanzi
ubuchwepheshe be-eto sterilization uma buqhathaniswa nabanye: Izinzuzo kanye Nobubi Ukuvala inzalo kuyinqubo ebalulekile ekulawuleni ukutheleleka, iqinisekisa ukuthi amathuluzi ezokwelapha
Ukuqhathanisa inzalo ye-ETO Nezinye Izindlela Funda kabanzi
Ibanga lokusetha izinga lokushisa le-ETO Sterilization Chamber kufanele libe phakathi kuka-30-60°C. Kunconywa ukwenza ukuhlolwa kokusatshalaliswa kwezinga lokushisa kwendawo yangaphakathi yesibulala-magciwane kanye nokuhlola izinga lokushisa kanye nomswakama wokusabalalisa ngaphakathi kwegumbi lokuvala inzalo ukuze kuqinisekiswe ukusebenza kahle kwezinhlelo zokushisisa nokuswakama.
I-ETO Sterilization Chamber Testing Funda kabanzi
Inhloso yokuqapha nokulawula ukuvalwa kwe-EO (Ethylene Oxide) ukukhombisa ukuthi inqubo eqinisekisiwe neshiwo yokuvala inzalo isetshenzisiwe emkhiqizweni. Idatha evela kumjikelezo ngamunye wokuvala inzalo kufanele irekhodwe futhi igcinwe ukuze kuqinisekiswe ukuthi ukucaciswa kwenqubo yokuvala inzalo okuqinisekisiwe kuhlangatsheziwe. Lokhu kufanele kufake okungenani okuqukethwe okulandelayo:
Ukuqapha Okujwayelekile, Ukulawula, kanye Nokukhishwa Kwe-Ethylene Oxide Sterilization Funda kabanzi
I-Process Challenge Device (PCD) iwuhlelo lwenselelo ye-microbiological esetshenziselwa ukuhlola izinga lokubulala lamapharamitha enqubo akhethiwe. Ngokuvamile idivayisi noma iphakheji yokuhlola equkethe izinkomba zebhayoloji (isb., izinhlamvu ze-Bacillus atrophaeus ezinesibalo se-spore esaziwayo). Ukumelana ne-PCD enqubweni yokuvala inzalo kufanele kube kukhulu noma kulingane nokumelana nomthwalo wemvelo we-microbial endaweni okunzima kakhulu ukuyibulala inzalo yomkhiqizo. Inani lama-PCD kufanele lihlangabezane nezidingo ezivezwe kuThebula C.3 leSithasiselo C kokuthi GB18279.1-2015. Ama-PCD ahlukaniswa ama-PCD angaphakathi (IPCD) nama-PCD angaphandle (EPCD).
Ukukhiqizwa Kwamadivayisi Wenselele Yenqubo ku-eo inzalo Funda kabanzi
Ukuqinisekiswa Kokufakwa Kwemboni ye-Eto Sterilizer Ukufakwa kwe-Industrial Eto Sterilizer kanye nazo zonke izinsizakalo ezihambisanayo kufanele kuhambisane nezidingo.
Ukuqinisekiswa Kokufakwa Kwemboni ye-Eto Sterilizer Funda kabanzi
Amanzi okuthambisa asetshenziswa ku-ETO Sterilization Equipment kufanele angabi nakungcola, akufanele enze buthaka inqubo yokuvala inzalo ye-ETO, futhi akumele alimaze isibulala-magciwane, imikhiqizo evalwe inzalo, noma umthwalo. I-Humidification kufanele yenziwe ngokujova isitimu, hhayi ngokujova ngokuqondile amanzi ane-athomu.
Izidingo Zamanzi Ahlanziwe Zokwenziwa Kwenzalo ye-ETO Funda kabanzi