ፓራሜትሪክ መልቀቅ እና የጸዳ ምርት መለቀቅ ፕሮቶኮሎች
ለፓራሜትሪክ መልቀቂያ ተገዢነት ከፍተኛ ትክክለኛነት EO የማምከን ክፍል።
1. ለፓራሜትሪክ መልቀቂያ የመከታተል ችሎታዎች
- የአየር ፍሰት ክትትል; በቅድመ ማቀዝቀዣ, በማምከን እና በአየር ማስወጫ ክፍሎች ውስጥ የአየር ዝውውር መሳሪያዎችን እና አናሞሜትሮችን አስገዳጅ መጫን.
- ቀጥተኛ እርጥበት; በሕክምናው ወቅት የእውነተኛ ጊዜ መለኪያ.
- የኢኦ ማጎሪያ፡ በተጋለጡበት ጊዜ በሁሉም አስፈላጊ ክፍተቶች ቀጥተኛ መለኪያ.
- ድርብ የሙቀት ነጥቦች; የክፍሉን ሁኔታ የሚወክሉ ቢያንስ ሁለት ገለልተኛ ዳሳሾች።
1.1 ዳሳሽ ቴክኖሎጂዎች & መለካት
የተቀናጀ የእርጥበት መጠን እና የኢኦ ማጎሪያ ቁጥጥር ስርዓቶች።
2. የምርት መለቀቅ (ምዕራፍ 11)
ለባች መልቀቅ የሰነድ ግምገማ እና የአካላዊ መለኪያ ቁጥጥር።
3. ልዩነቶች እና ምርመራዎች
ማስተካከያ እና ጥገና ለቀጣይ የፓራሜትሪክ መልቀቅ የዳሳሽ ትክክለኛነትን ማረጋገጥ።
Key figures
| መለኪያ | መስፈርት | Reference |
|---|---|---|
| የሙቀት መጠን | Continuous measurement by a calibrated sensor; the value used for release is the recorded value, not the setpoint | ጂቢ / ዜድ 44877-2024 |
| Relative humidity | Measured directly in the chamber rather than calculated from the load | ጂቢ / ዜድ 44877-2024 |
| EO concentration | Determined by spectroscopy or gas chromatography, with accuracy and precision documented | ጂቢ 18279-2023 |
| Pressure | Logged through the vacuum, exposure and wash-out phases | ጂቢ 18279-2023 |
| መለካት | Traceable, on a defined interval, covering temperature, humidity and pressure loops | ISO 11135 |
| Batch release | Based on the recorded cycle parameters; biological indicators are not required for routine release | ጂቢ / ዜድ 44877-2024 |
በተደጋጋሚ የሚጠየቁ ጥያቄዎች
Which parameters must be monitored continuously for parametric release?
Airflow in the preconditioning, sterilization and aeration chambers, direct humidity during treatment, EO concentration at every required interval of exposure, and at least two independent temperature points representing the chamber state.
How are humidity and EO concentration measured directly?
Humidity by capacitance film for relative humidity or spectroscopy for absolute humidity, with sensors calibrated against saturated salt solutions. EO concentration by infrared spectroscopy or gas chromatography, using heat-traced sampling lines to prevent condensation.
Can a batch be released without biological indicators?
Under parametric release, yes — release rests on the physical parameters meeting their validated specifications, not on BI results. A documented review by authorised personnel is still mandatory before the batch is released.
What happens when a cycle deviates from its specification?
It cannot be released on the strength of a passing BI. A process that fails its parametric specifications must be investigated and dispositioned; parametric release can only resume after repair and re-calibration of the equipment.
Which standards govern parametric release?
GB/Z 44877-2024 and GB 18279-2023 for the Chinese market, which set the monitoring, sensor calibration and documented-review requirements the equipment has to meet.