1.3 30 m³ eto gas sterilizer Komposisyon

Ang 30 m³ eto gas sterilizer ay binubuo ng sterilizer housing, auxiliary equipment (heating at thermal cycling device, vacuum extraction device, dosing at vaporization device, humidification device, sealing device, natitirang gas processing device), at monitoring at control device.

1.4 Pag-uuri

Batay sa laki ng eto gas sterilizer, may dalawang uri: Class A at Class B.

  • Class A eto gas sterilizer — User-programmable sterilizer na may volume ng sterilization chamber > 1 m³, angkop para sa isterilisasyon sa industriya ng pagmamanupaktura ng medikal na aparato.
  • Class Beto gas sterilizer — Isang sterilizer na may isa o higit pang preset na mga siklo ng isterilisasyon at mga limitasyon sa laki, na may volume ng sterilization chamber ? 1 m³, angkop para sa pag-sterilize ng mga klinikal na kagamitang medikal.

1.5 Talahanayan ng Pagkakaiba ng Modelo

Talahanayan 1: Mga Modelo at Pangunahing Parameter ng Ethylene Oxide Sterilizer

Serial No.ModeloMga Dimensyon ng Panloob na Kamara (L×W×H) (mm) ±1%Nominal na Dami (m³)
Power (kW)Timbang (kg)Power Supply
1BCS-11400*700*10001m³161000380V/50Hz
2BCS-21650*1000*12002m³162000380V/50Hz
3BCS-31400*1350*17003.0m³163000380V/50Hz
4BCS-4.52000*1350*17004.5m³274000380V/50Hz
5BCS-52250*1350*17005m³274500380V/50Hz
6BCS-62700*1350*17006m³275000380V/50Hz
7BCS-83500*1350*17008m³276000380V/50Hz
8BCS-104500*1350*170010m³437000380V/50Hz
9BCS-155500*1350*200015m³569000380V/50Hz
10BCS-207500*1350*200020m³8112000380V/50Hz
11BCS-269600*1350*200026m³8114000380V/50Hz
12BCS-2812500*1350*170028m³9014000380V/50Hz
13BCS-3011300*1350*200030m³9017000380V/50Hz
14BCS-30-25600*2700*200030m³-29016000380V/50Hz
15BCS-3513000*1350*200035m³10023000380V/50Hz
16BCS-35-26000*2700*220035m³-210022000380V/50Hz
17BCS-40-26650*2700*220040m³-210023000380V/50Hz
18BCS-4812850*1400*270048m³10028000380V/50Hz
19BCS-5213900*1400*270052m³15028000380V/50Hz
20BCS-55-27500*2700*270055m³-215027000380V/50Hz
21BCS-57-29650*2700*220057m³-215028000380V/50Hz
22BCS-63-28650*2700*270063m³-215029000380V/50Hz
23BCS-70-29700*2700*270070m³-215532000380V/50Hz
24BCS-78-210750*2700*270078m³-216535000380V/50Hz
25BCS-86-211800*2700*270086m³-216538000380V/50Hz
26BCS-101-217050*2700*2200101m³-220045000380V/50Hz
27BCS-120-220200*2700*2200120m³-220049000380V/50Hz

Sterilization Cycle and Automatic Control

2.10.7.1

The 30 m³ eto gas sterilizer should be equipped with an automatic controller that can automatically control the sterilization cycle.

2.10.7.2 Class A eto gas sterilizer:

  • The automatic controller should be able to program preset conditions for each stage of the sterilization cycle.
  • During the sterilization cycle, it must not change the preset variables.

2.10.7.3 Class B eto gas sterilizer:

  • The manufacturer applies the automatic controller to program the preset conditions for each stage of the sterilization cycle.
  • The manufacturer should provide methods to shorten the sterilization time for verification purposes, which can only be performed using a special key, password, or tool. Once the sterilization time is shortened, there should be a fault indication at the end of the operating cycle.

2.10.7.4

The controller should only be accessible using a special key, password, or tool. Additionally: a) For Class A sterilizers, access to the controller and setting of operational cycle variables should require a special key, password, or tool.
b) For Class B sterilizers, only the manufacturer can program new operational cycles.

2.10.7.5

Sa panahon ng pagpapanatili, pagsubok, o emerhensiya, ang programa ng awtomatikong controller ay dapat na manual na gumagana. Ang manual na device na ginamit ay dapat na iba sa espesyal na key, password, o tool na tinukoy sa seksyon 2.10.7.4. Kapag isinagawa ang manu-manong operasyon dahil sa pagpapanatili, pagsubok, o mga emerhensiya, ang awtomatikong programa ay dapat na awtomatikong huminto. Ang sistema ng manu-manong pagpapatakbo ay hindi dapat magdulot ng banta sa kaligtasan, na nagpapahintulot lamang sa isang yugto na piliin nang sunud-sunod sa isang pagkakataon.

2.10.7.6

Sa matagumpay na pagkumpleto ng bawat tinukoy na ikot ng isterilisasyon, dapat ipahiwatig ng awtomatikong controller na kumpleto na ang ikot.