Process Water For ETO Sterilization Equipment

ETO 滅菌装置で使用される加湿水には汚染物質が含まれておらず、ETO 滅菌プロセスを弱めたり、滅菌器、滅菌製品、または負荷に損傷を与えたりしてはなりません。加湿は霧化した水を直接噴射するのではなく、スチームを噴射して行う必要があります。
Note: It is recommended to use purified water that meets the requirements of the Chinese Pharmacopoeia for humidification. If purified water is not used, the steam supply or steam condensate should be validated to ensure that the humidification steam meets the requirements of this section. The steam quality testing methods are outlined in YY/T 1612 and the Chinese Pharmacopoeia.
Steam Contaminant Limits
| Contaminant | Condensate | Steam Supply |
|---|---|---|
| Residue | Less than 10mg/kg | Less than 100mg/L |
| Heavy Metals | Not exceeding the limits for purified water in the Chinese Pharmacopoeia | Not exceeding the limits for purified water in the Chinese Pharmacopoeia |
| Chlorides | Less than 1mg/kg | Less than 10mg/L |
| Conductivity | Less than 15?S/cm | Less than 50?S/cm |
| pH | 5–8.5 | 5–8.5 |
| Appearance | Colorless | Colorless |
Note 1: If the contaminant levels in the feedwater or steam exceed the above limits, the sterilizer’s service life will be significantly reduced.
Process Air For Sterilization
The air used in the sterilizer should be dried and filtered to ensure the service life and filtering performance of the ETO Sterilization sterilizer’s terminal filter.
さらに詳しく知りたい場合は
電子メール:ヘイリー@hzbocon.com
ローカルサイト: 中国浙江省杭州市下沙区財通中新1202号室
Webサイト:hzbocon.comzjbocon.com