Process Water For ETO Sterilization Equipment

ETO 滅菌装置で使用される加湿水には汚染物質が含まれておらず、ETO 滅菌プロセスを弱めたり、滅菌器、滅菌製品、または負荷に損傷を与えたりしてはなりません。加湿は霧化した水を直接噴射するのではなく、スチームを噴射して行う必要があります。

Note: It is recommended to use purified water that meets the requirements of the Chinese Pharmacopoeia for humidification. If purified water is not used, the steam supply or steam condensate should be validated to ensure that the humidification steam meets the requirements of this section. The steam quality testing methods are outlined in YY/T 1612 and the Chinese Pharmacopoeia.

Steam Contaminant Limits

ContaminantCondensateSteam Supply
ResidueLess than 10mg/kgLess than 100mg/L
Heavy MetalsNot exceeding the limits for purified water in the Chinese PharmacopoeiaNot exceeding the limits for purified water in the Chinese Pharmacopoeia
ChloridesLess than 1mg/kgLess than 10mg/L
ConductivityLess than 15?S/cmLess than 50?S/cm
pH5–8.55–8.5
AppearanceColorlessColorless

Note 1: If the contaminant levels in the feedwater or steam exceed the above limits, the sterilizer’s service life will be significantly reduced.

Process Air For Sterilization

The air used in the sterilizer should be dried and filtered to ensure the service life and filtering performance of the ETO Sterilization sterilizer’s terminal filter.

さらに詳しく知りたい場合は

電話:+8619975258603

電子メール:ヘイリー@hzbocon.com

ローカルサイト: 中国浙江省杭州市下沙区財通中新1202号室