5.1 Normal Operating Conditions

5.1.1 The normal working environment for the sterilizer should meet the following conditions:

a) Ambient temperature: 5 ? to 40 ?;
b) Relative humidity: ? 80 % RH;
c) Atmospheric pressure: 70 kPa to 106 kPa;

5.1.2 Rated Operating Conditions of the Sterilizer

The rated operating conditions for each model of the sterilizer are shown in Table 2.

МодельSupply VoltageFrequencyPower (kW)Current (A)
БКС-1380V50Hz1624
БКС-21624
БКС-31624
БКС-4,52741
БКС-52741
БКС-62741
БКС-82741
БКС-104365
БКС-155685
БКС-2081123
БКС-2681123
БКС-2890137
БКС-3090137
БКС-30-290137
БКС-35100152
БКС-35-2100152
БКС-40-2100152
БКС-48100152
БКС-52150228
БКС-55-2150228
БКС-57-2150228
БКС-63-2150228
БКС-70-2155236
БКС-78-2165251
БКС-86-2165251
БКС-101-2200304
БКС-120-2200304
  • The power supply socket is a three-phase 380V, three-hole socket with a grounding wire.
  • The power plug is a three-phase 380V, three-prong plug with a grounding wire.
  • A yellow-green outer wire is used to connect to the protective conductor terminal.

5.2 Storage Environment of the 12m3 eto sterilizer

The sterilizer should be stored in an indoor environment with temperatures between -20? and +55?, relative humidity ? 80% RH, free from corrosive gases, and well-ventilated.
Примечание: Before storage, all water in the circulating water system must be drained to keep the system in a dry state.

5.3 Transportation Conditions of the 12m3 eto sterilizer

The sterilizer should be transported unpackaged, and when hoisting the sterilizer, it should be lifted using the device’s own lifting rings. During transport by vehicle, it should be secured as per regulations.

5.4 Disposal of Decommissioned 12m3 Sterilizers

After decommissioning, the sterilizer should be disposed of by a qualified recycling company.