eto sterilization process

eto sterilization process

Implementation of the ETO Sterilization Process

Eto Sterilization is conducted using a eto sterilizer located in a dedicated facility. Outside the sterilization room, control equipment regulates the process according to established parameters and requirements.

Composition of Agents Used in the ETO Sterilization Process

The agents employed in sterilization consist of either pure ethylene oxide or mixtures of carbon dioxide and nitrogen.

Precautionary Measures for Ethylene Oxide

Ethylene oxide (EO) gas must be handled, stored, and used with extreme caution due to its highly toxic nature, flammability, and explosion risk at concentrations as low as 2.6% in air.

Steps Involved in the Sterilization Process

The eto sterilization process comprises two critical steps:

  1. Preconditioning
  2. Sterilization Process

Preconditioning

Preconditioning involves achieving the necessary humidity and temperature levels prior to introducing the load into the chamber, ensuring effective sterilization unaffected by internal environmental factors.

Sterilization Cycle

The EO sterilization process includes the following elements:

1.加熱
2.保溫
3.預真空
4.洩漏測試
5.加濕
6.濕度住宅
7.EO注射
8. EO注射液2
9.EO注射液3
10.EO住宅
11.EO去除2
12.空氣清洗
13.曝氣

Influence of Load Chamber

The configuration of the load within the chamber affects:

  • Heat distribution patterns
  • Humidity levels
  • EO gas penetration
  • EO gas removal

Design of Load Configuration

The load configuration should be designed to facilitate the sterilization process by:

  • Removing air
  • Aiding heat penetration
  • Eliminating humidity
  • Ensuring EO removal during and after the process

Process Characterization & Safe Delivery of Sterilization Processes

Process characterization defines the range of variables required by ISO 11135 and identifies four key process variables, which must be documented at every stage of the sterilization process:

  1. Phase
    Covers variables associated with each phase of the process.

  2. Pre-conditioning
    Involves time, temperature, humidity, and transfer time.

  3. Sterilization Cycle
    Includes exposure time, temperature, humidity, EO concentration, and pressure.

  4. Aeration
    Pertains to time and temperature during the aeration or gassing phase.

發表留言