9.1 Ngesikhathi sokwenziwa kwe-ETO Sterilization, abaqhubi abavunyelwe ukuphuma endaweni yokusebenza futhi kufanele babheke ngokuqhubekayo inqubo ye-ETO Sterilization. Uma kwenzeka noma yiziphi izinto ezingavamile, umshini kufanele uvalwe ngokushesha, futhi ugesi kufanele unqanyulwe ukuze uhlolwe.

9.2 Shintsha amanzi ajikelezayo ohlelweni lokushisa njalo ezinyangeni ezintathu. Lapho ushintsha amanzi, qala uvule wonke amaswishi okukhipha amanzi ngaphansi kwegumbi kanye nengxenye engezansi yomnyango ukukhipha amanzi ngokuphelele. Ngemva kokugeza ithange lamapayipi nelamanzi ngamanzi ahlanzekile, phinda ugcwalise ngamanzi amasha ajikelezayo.

9.3 Hlola futhi uhlanze umoya, ukufudumeza, ukuhwamuka, kanye nezihlungi zesistimu ye-ethylene oxide okungenani ngemva kwayo yonke imijikelezo engama-30 ye-ETO Sterilization, esikhundleni sanoma yiziphi izihlungi ezilimele noma ezingasebenzi kahle. (Izindlela zokuhlanza zingabandakanya amanzi, umoya ocindezelweyo, noma isitimu.)

9.4 Amageji e-vacuum pressure, izinzwa zomfutho, nezinye izinsimbi zokulinganisa kufanele zilinganiswe umnyango wendawo we-metrology njalo ezinyangeni eziyisithupha, ngokuya ngemithetho kazwelonke ye-metrology.

9.5 Ama-gaskets okuvala kufanele ashintshwe njalo ezinyangeni eziyisithupha.

9.6 Njalo hlola ukuthi zonke izixhumi ezixhumene nogesi ohlelweni lokulawula ugesi zivikelekile.

9.7 Okungenani kanye ngosuku, hlola amapayipi okujikeleza kwamanzi, ivacuum, i-ethylene oxide, isitimu, kanye namapayipi edivayisi yokuhwamuka ukuze avuze. Uma kukhona ukuvuza okutholakalayo, kufanele kuthathwe izinyathelo zokubhekana nakho ngokuphepha.

9.8 Gcoba izinti zezicabha kanye ngesonto futhi uhlanze izindonga zangaphakathi, uhlaka lomnyango, namahinji ukuze uqinisekise ukuthi akukho kungcola okunqwabelanayo ezindongeni zangaphakathi noma amahinji.

9.9 Uma izinga lokushisa le-ambient elizungeze okokusebenza lehla ngaphansi kwama-degree Celsius angu-zero futhi okokusebenza kungakwazi ukusebenza ngokuqhubekayo, uma isimiso sokushisisa siyekile ukushisisa isikhathi esingaphezu kwamahora amathathu, amanzi ohlelweni lokujikeleza kufanele akhishwe ukuze kuvinjwe isistimu ekuqandiseni nasekuqhekekeni.