Twenty practical questions answered by sterilization engineers — from which products are suitable for ethylene oxide, to how validation cycles work, packaging material selection, bioburden testing, and sterile shelf life.

Érdeklődni WhatsApp-on keresztül View ETO Sterilizer Product

Ethylene oxide (EO) sterilization is a rigorously controlled special process used across medical devices, pharmaceuticals, and industrial products. Because its efficacy cannot be verified by post-sterilization testing alone, every product must undergo a formal validation before routine processing. The following guide answers the most common questions we receive from manufacturers preparing for ETO sterilizálás.

Which Products Are Suitable for ETO Sterilization?

Medical Devices

Széles körben használják műanyag, fém és nem szőtt orvosi eszközökhöz, beleértve az infúziós szerelékeket, fecskendőket, biopsziás fogókat, orvosi katétereket, sebészeti köpenyeket, maszkokat és sebészeti bilincseket.

Ipari & Egyéb alkalmazások

Alkalmas bőrhöz, pamuttermékekhez, szintetikus szövetekhez, precíziós műszerekhez, biológiai termékekhez, papírhoz, könyvekhez, dokumentumokhoz, bizonyos gyógyszerekhez és gumitermékekhez – minden olyan hőérzékeny dologhoz, amely más módszerrel nem sterilizálható.

Felkészülés a sterilizálásra

A termékem a gyártás után azonnal sterilizálható?

No. ETO sterilization efficacy cannot be judged by post-sterilization sterility testing alone — it is a special process that requires strict control of the entire procedure. This control comes from careful sterilization parameter design and confirmed validation results. Using an unvalidated sterilization process risks inadequate sterilization, which could lead to bacterial infection in patients. For this reason, regulations in most countries mandate sterilization validation before routine processing.

What exactly is sterilization validation?

In simple terms, validation means designing the sterilization process and its parameters to achieve a predetermined sterilization effect, then running multiple experiments to obtain reliable process parameters that guarantee the outcome. These experiments serve two purposes: (1) confirming that the product’s sterility assurance level (SAL) meets the target; (2) establishing routine sterilization parameters and forming documented operational and control requirements.

I am not sure what chamber size to choose for sterilization.

The chamber size mainly depends on your sterilization volume and delivery cycle requirements. Larger chambers offer higher throughput but require more floor space and utilities. We recommend evaluating your projected annual volume, batch size, and turnaround time before selecting a model.

Sterility Assurance Level (SAL)

What is SAL? Sterility assurance level is the probability of detecting a viable microorganism on a single product unit after sterilization. It is typically expressed as 10-6 or 10-3. When used for SAL, 10-6 represents a lower (better) probability than 10-3 — meaning the product is safer. Regulatory experts worldwide consider a probability of 10-6 to be safe based on risk analysis.

Can I simply perform sterility testing after routine sterilization instead of validation?

Nem. Tegyük fel, hogy egy sterilizációs tétel 100 dobozt tartalmaz, egyenként 100 egységgel – összesen 10 000 egység. A GB2828 alapján a legmagasabb AQL határ 0,01, ami azt jelenti, hogy a mintavételezés csak 1/10 000 minőségi szintet garantál. A 10-6 a követelmény alapvetően nem észlelhető. Még a 0,01-es szinten is 1250 mintát [0 1] kell venni, és mindegyiknek sterilnek kell lennie ahhoz, hogy a 10 000 egységet elfogadhatónak ítélje meg. A költség és a kockázat megfizethetetlen. Maga a sterilitásteszt körülbelül 1/100-as hamis pozitív arányt mutat. Mind elméletben, mind gyakorlatban a sterilizálás utáni sterilitásvizsgálat nem helyettesítheti a validálást.

Érvényesítési idővonal & Ciklusok

Mennyi ideig tart a sterilizálás validálása?

At least 45 days. A first-time validation typically runs 7 cycles: 4 half cycles, 1 short cycle, and 2 full cycles. Each half cycle and short cycle requires pre-cooling the product for 48 hours to bring it down to 0°C, then running preconditioning, sterilization, and aeration for approximately 24 hours. Including preparation of BIs and sensors, one cycle takes about 4 days, so 7 cycles take roughly 28 days. Adding testing time and report preparation, 45 days is the minimum. Deviations or unexpected results require re-adjusting parameters and running additional cycles, and since the sterilizer also handles routine production, actual validation often takes 60–90 days. Complex products may require even more time.

What is the purpose of the half cycle in validation?

The half cycle is one of the methods specified in ISO 11135 to demonstrate SAL = 10-6. With all sterilization parameters unchanged except exposure time, it determines the shortest EO exposure time that produces no surviving microorganisms. Doubling that time under the same conditions gives the full cycle, which is expected to achieve SAL = 10-6. Three replicate runs are required by the standard.

What is the purpose of the short cycle in validation?

The short cycle evaluates the resistance relationship between IPCD, EPCD, and the product itself. Through the short cycle, the requirement is EPCD resistance ? IPCD resistance ? product resistance. The product must pass all sterility tests (recommended per pharmacopeia requirements).

What is the purpose of the full cycle in validation?

A teljes ciklus meghatározza a sterilizálási folyamat felső paraméterhatárait, megkérdőjelezve a termék fizikai és kémiai tulajdonságait (beleértve a sterilizálószer-maradékot is). A sterilizálás utáni minták egyikét egy második sterilizálási ciklusnak vetik alá, hogy megkérdőjelezzék a kettős sterilizálás megbízhatóságát. A termékteljesítmény-tesztelemeket, mennyiségeket és módszereket a vásárló határozza meg. Az EO/ECH-t gázkromatográfiával is tesztelhetjük, de nem vagyunk akkreditált laboratórium. Ha akkreditált tesztelésre van szükség, akkor azt ki kell szervezni. Az EO/ECH vizsgálatnak időbeli korlátai vannak – a kihelyezett mintákat szárazjégen kell szállítani, és a laboratóriumnak szárazjégen kell tartania a csomagot, amikor kinyitja.

Bioterhelés tesztelése

Is bioburden testing required for validation samples?

Yes. All sterile medical devices should have their bioburden tested periodically. Validation samples must be accompanied by a bioburden test report from product manufactured under normal production conditions. This work is performed by the product manufacturer.

Must bioburden samples be produced in a cleanroom?

Bioburden test samples must come from the normal production process — the exact same process as products that are routinely sterilized. All samples used for actual testing should be produced through a process identical to routine sterilization production.

Packaging Considerations

My product is packaged in plastic (PE film). Can it be sterilized with EO?

Az ISO 11135 szabvány nem engedélyezi a műanyag zacskóba csomagolt termékek EO-sterilizálását, helyette papír-műanyag dializáló tasakot ajánl. Egyes országokban azonban nincs erre vonatkozó kifejezett szabályozás, és a gyakorlatban a műanyag zacskóba csomagolt termékeket továbbra is EO-sterilizálják, speciálisan konfigurált paraméterekkel. Műanyag zacskós csomagolás esetén a levegőztetés nagyon nehéz, gyakran tíz vagy több napot vesz igénybe. Javasoljuk, hogy szellőztetőképes csomagolóanyagokat használjon.

A termékem nem hőálló. Lehet EO-val sterilizálni?

Our sterilization parameters generally operate at 40–60°C. We recommend that materials, packaging, and process auxiliaries withstand temperatures above 60°C to avoid product damage. If the product has special temperature limitations, low-temperature sterilization must be specially validated.

After EO sterilization, some bonded areas have come loose or deformed (bending, corner lifting, etc.). What causes this?

EO sterilization requires specific temperatures, typically around 54°C. Some products have poor heat resistance, and temperatures in the 50s can reduce physical properties such as bond strength, causing these phenomena. This is exactly why product sterilization validation is essential — to confirm a suitable sterilization method and parameters under which the product achieves the required SAL while still maintaining its intended characteristics.

Routine Sterilization & Release

Does routine sterilization require product sterility testing?

Nem feltétlenül. Az ISO szabványok szerint a termék sterilitásának vizsgálata általában nem szükséges a rutinszerű kiadáshoz. Mindaddig, amíg a validálást az ISO 11135 szerint hajtják végre, és a rutin sterilizálás a szabványt követi, az aszeptikus kibocsátás elfogadható. Kínában azonban a steril termék kibocsátásának meg kell felelnie a termékregisztrációs szabványnak is – azt a szabványt, amelyet a vállalat az NMPA-nál való regisztrációs tanúsítványának kérelmezésekor hozott létre. Sok vállalat a regisztrációs szabvány elkészítésekor a régi megközelítést követte, amely szerint a tételenkénti sterilitási vizsgálatot megkövetelték. Mivel a cég maga határozta meg, ezért teljesíteni kell.

A rutin sterilizáláshoz szükséges az EO/ECH maradékanyag vizsgálata?

Nem feltétlenül. Az ISO szabványok szerint az EO/ECH rutinszerű maradékanyag-vizsgálatára általában nincs szükség. Mindaddig, amíg a maradékanyag-kibocsátás ellenőrzése az ISO 10993-7 szerint történik, és a rutin levegőztetés követi a szabványt, ez elfogadható. Kínában azonban a steril termék kibocsátásának is meg kell felelnie a termékregisztrációs szabványnak. Sok vállalat kötegenkénti EO/ECH maradékvizsgálatot írt elő regisztrációs szabványában, ezért azt el kell végezni.

Az érvényesítési jelentés elkészülte után mennyi időn belül kezdődhet meg a sterilizálás?

Once the validation report is complete, the customer must approve it. Sterilization can begin immediately after approval. If sterilization is urgent, products can be shipped first and sterilized, but they cannot be released until both parties agree on the validation results and all post-sterilization documentation fully meets the specified requirements.

Sterile Shelf Life & Post-Sterilization Issues

How long can my product be stored after EO sterilization?

EO sterilization is a microbial killing process. If the environment after sterilization remains sterile, the product remains sterile indefinitely — as long as the environment does not change. In reality, the sterile shelf life of a product has little to do with the EO sterilization process itself. Maintaining sterile life depends on the primary packaging (also called sterile packaging or sterile barrier). Tyvek material typically maintains sterility for over 5 years, while ordinary medical dialysis paper maintains around 2 years. The shelf life of sterile packaging must be challenged — refer to ISO 11607 and ASTM 1989. Note that both EO and irradiation sterilization challenge the sterile packaging system. If the packaging system is not strong enough, it may be damaged after sterilization, and the product will be contaminated immediately after sterilization.

After sterilization, my customer’s testing found the product non-sterile. Does this mean sterilization failed?

A non-sterile test result can have many causes: (1) false positive — a retest is usually needed to confirm; (2) weak pouch seal or poor microbial barrier of the material; (3) sterilization parameter deviation; (4) load pattern change; (5) packaging material change; (6) transportation damage compromising seal integrity; (7) bioburden deviation, such as unusually high counts with increased resistance; (8) product structure trapping microorganisms internally, while testing uses a cutting method — EO sterilization is a surface sterilization process.

After sterilization, the color-change label (chemical indicator) shows uneven or incomplete color change. Is this a sterilization failure?

Uneven color change on the indicator label has many causes and is completely different from sterilization failure. Chemical indicators respond to specific conditions such as temperature, humidity, and EO concentration; uneven color change often reflects uneven distribution of those conditions within the load rather than inadequate lethality. Proper interpretation requires reviewing the full cycle records and biological indicator results.

Have More Questions About ETO Sterilization?

Mérnöki csapatunk segíthet a termék alkalmasságának értékelésében, a tervezési érvényesítési protokollokban, a csomagolóanyagok kiválasztásában és a sterilizálási munkafolyamat megtervezésében. Lépjen kapcsolatba velünk, hogy megbeszéljük konkrét igényeit.