5.1 Normal Operating Conditions
5.1.1 The normal working environment for the sterilizer should meet the following conditions:
a) Ambient temperature: 5 ? to 40 ?;
b) Relative humidity: ? 80 % RH;
c) Atmospheric pressure: 70 kPa to 106 kPa;
5.1.2 Rated Operating Conditions of the Sterilizer
The rated operating conditions for each model of the sterilizer are shown in Table 2.
Model | Supply Voltage | Frequency | Power (kW) | Current (A) |
---|---|---|---|---|
BCS-1 | 380V | 50Hz | 16 | 24 |
BCS-2 | 16 | 24 | ||
BCS-3 | 16 | 24 | ||
BCS-4.5 | 27 | 41 | ||
BCS-5 | 27 | 41 | ||
BCS-6 | 27 | 41 | ||
BCS-8 | 27 | 41 | ||
BCS-10 | 43 | 65 | ||
BCS-15 | 56 | 85 | ||
BCS-20 | 81 | 123 | ||
BCS-26 | 81 | 123 | ||
BCS-28 | 90 | 137 | ||
BCS-30 | 90 | 137 | ||
BCS-30-2 | 90 | 137 | ||
BCS-35 | 100 | 152 | ||
BCS-35-2 | 100 | 152 | ||
BCS-40-2 | 100 | 152 | ||
BCS-48 | 100 | 152 | ||
BCS-52 | 150 | 228 | ||
BCS-55-2 | 150 | 228 | ||
BCS-57-2 | 150 | 228 | ||
BCS-63-2 | 150 | 228 | ||
BCS-70-2 | 155 | 236 | ||
BCS-78-2 | 165 | 251 | ||
BCS-86-2 | 165 | 251 | ||
BCS-101-2 | 200 | 304 | ||
BCS-120-2 | 200 | 304 |
- The power supply socket is a three-phase 380V, three-hole socket with a grounding wire.
- The power plug is a three-phase 380V, three-prong plug with a grounding wire.
- A yellow-green outer wire is used to connect to the protective conductor terminal.
5.2 Storage Environment of the 12m3 eto sterilizer
The sterilizer should be stored in an indoor environment with temperatures between -20? and +55?, relative humidity ? 80% RH, free from corrosive gases, and well-ventilated.
Note: Before storage, all water in the circulating water system must be drained to keep the system in a dry state.
5.3 Transportation Conditions of the 12m3 eto sterilizer
The sterilizer should be transported unpackaged, and when hoisting the sterilizer, it should be lifted using the device’s own lifting rings. During transport by vehicle, it should be secured as per regulations.
5.4 Disposal of Decommissioned 12m3 Sterilizers
After decommissioning, the sterilizer should be disposed of by a qualified recycling company.