EO sterilization equipment and process characteristic patterns and effect analysis

Start cycle phase(Select parameters/sterilization recipe)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Choosing the wrong recipeLack of care in choosing the formulationInvalid cycle. Affects product sterility (re-sterilization)1. Different recipe names65103001. Work instructions 2. Careful inspection before the cycle starts 3. Operator training254401. Work instructions 2. Careful inspection before starting the cycle 3. Operator traininguser
Loss of product/packaging functionality and integrity due to unvalidated cycle parameters1. Different recipe names64102401. Work instructions 2. Careful inspection before the cycle starts 3. Operator training244321. Work instructions 2. Careful inspection before starting the cycle 3. Operator traininguser
Warm-up phase(Cabinet heating)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation532301. Preventive maintenance232121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Automatic heating time compensation632361. Preventive maintenance232121. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)
1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm
652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
2The temperature inside the cabinet is too highTemperature sensor failureAffects product sterility (re-sterilization)
1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm
642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
Insulation stage(Maintain cabinet temperature)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm532301. Preventive maintenance232121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm632361. Preventive maintenance232121. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Heating timeout alarm 2. Circulating water pump overload alarm652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
2The temperature inside the cabinet is too highTemperature sensor failureAffects product sterility (re-sterilization)1. Heating timeout alarm 2. Circulating water pump overload alarm642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
3Lack of timeProgramming ErrorsRepeat sterilization1. Automatic cycle inspection24216none
Pre-vacuum stage(Removing air from the chamber allows the process to proceed below the EO flammability limit)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Not enough air removedLow vacuum pump efficiencyThe cycle is locked in one phase until it is restored and restarted after a standby period1. Pre-vacuum time over limit alarm 2. Vacuum pump overload alarm 3. Automatically detect water flow and shut down the waterless pump532301. Preventive maintenance232121. Preventive maintenanceuser
Wrong valveThe cycle is locked in one phase until it is restored and restarted after a standby period1. Pre-vacuum time over limit alarm 2. Vacuum pump overload alarm632361. Preventive maintenance232121. Preventive maintenanceuser
Pressure sensor failureFlammable Zone Hazard1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit632361. Preventive maintenance 2. Calibration232121. Preventive maintenance 2. Calibrationuser
2Too much vacuumPressure sensor failureProduct and/or packaging damage (classification)1. Redundant pressure sensor 2. Pressure difference alarm is issued when the pressure difference is greater than 1kpa.642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
Leak detection stage(Check the tightness of the box)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Cabinet leakageGasket leakageNo impact1. Leak test failure alarm 2. Process ends10521001. Preventive maintenance352301. Preventive maintenanceuser
Valve leakageNo impact1. Leak test failure alarm632361. Preventive maintenance232121. Preventive maintenanceuser
Circulation system air duct leakageNo impact1. Leak test failure alarm432241. Preventive maintenance232121. Preventive maintenance 2. Calibrationuser
Nitrogen replacement stage(Fill the chamber with an inert gas to replace oxygen)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Not enough nitrogen entering the chamberInsufficient nitrogenThe cycle is locked in one phase until it is restored and restarted after a standby periodEnd of process33218N/A
Faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby periodN/A633541. Preventive maintenance233181. Preventive maintenanceuser
Pressure sensor failureFlammable Zone Hazard1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit632361. Preventive maintenance232121. Preventive maintenanceuser
2Too much nitrogen in the cabinet doorFaulty valveEO concentration is lower than the validated value because during EO injection, if controlled by pressure (resterilization), additional N2 is injected1. Indoor high voltage alarm 2. End of process6541201. Preventive maintenance254401. Preventive maintenanceuser
Pressure transmitter failureToo much EO gas injected. High EO residue. Extended degassing time me)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
3The cabinet temperature is higher than the upper limit tImpact on product function (sorting)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
4The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation632361. Preventive maintenance232121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Automatic heating time compensation633541. Preventive maintenance233181. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
Low carburetor temperatureAffects product sterility (re-sterilization)1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation642481. Preventive maintenance242161. Preventive maintenanceuser
Humidification stage(Increasing the humidity of the load by injecting steam)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Insufficient steamSteam generator failureAffects product sterility (re-sterilization)1. Alarm when steam pressure is lower than the limit652601. Preventive maintenance252201. Preventive maintenanceuser
Valve failureThe cycle is locked in one phase until it is restored and restarted after a standby period1.Humidification frequency limit alarm652601. Preventive maintenance252201. Preventive maintenanceuser
Pressure transmitter/humidity sensor failureAffects product sterility (re-sterilization)1. Redundant pressure/humidity sensor 2. Alarm when pressure/humidity difference exceeds limit652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
Filter cloggedThe cycle is locked in one phase until it is restored and restarted after a standby period1.Humidification frequency limit alarm652601. Preventive maintenance252201. Preventive maintenanceuser
2Too much steam injectionFaulty valveExcessive steam injection, wet cartons (sorting)1. Indoor high voltage alarm 2. End of process6541201. Preventive maintenance254401. Preventive maintenanceuser
Pressure transmitter/humidity sensor failureExcessive steam injection, wet cartons (sorting)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
3Temperature is too highTemperature sensor failureImpact on product functions (ranking)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
4The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked in one phase until it is restored and restarted after a standby period1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation632361. Preventive maintenance242161. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked in one phase until it is restored and restarted after a standby period1. Heating timeout alarm 2. Automatic heating time compensation633541. Preventive maintenance233181. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Heating timeout alarm 2. Circulating water pump overload alarm652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
Humidification dwell phase(Balanced humidity)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Lack of timetime outAffects product sterility (re-sterilization)1. Automatic cycle inspection252201. Preventive maintenance152101. Preventive maintenanceuser
Nitrogen cushion stage(Fill the chamber with inert gas to dilute the oxygen)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Not enough nitrogen entering the cabinetLack of nitrogenThe cycle is locked in one phase until it is restored and restarted after a standby period1. Check N2 pressure before starting the cycle 2. Alarm when N2 pressure is lower than the limit332180
Valve failureThe cycle is locked in one phase until it is restored and restarted after a standby periodnone633541. Preventive maintenance23318user1. Preventive maintenance
Pressure sensor failureFlammable Zone Hazard1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit632361. Preventive maintenance23212user1. Preventive maintenance
2Too much nitrogen injected into the cabinetFaulty valveEO concentration is lower than the validated value because during EO injection, if controlled by pressure (resterilization), additional N2 is injected1. Indoor high voltage alarm 2. End of process6541201. Preventive maintenance25440user1. Preventive maintenance
Pressure transmitter failureToo much EO gas injected. High EO residue. (Extend degassing time)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit652601. Preventive maintenance 2. Calibration25220user1. Preventive maintenance 2. Calibration
3The temperature inside the cabinet exceeds the upper limitTemperature sensor failureImpact on product function (sorting)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit642481. Preventive maintenance 2. Calibration24216user1. Preventive maintenance 2. Calibration
4The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation632361. Preventive maintenance23212user1. Preventive maintenance
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation633541. Preventive maintenance23318user1. Preventive maintenance
Temperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm652601. Preventive maintenance 2. Calibration25220user1. Preventive maintenance 2. Calibration
Low carburetor temperatureAffects product sterility (re-sterilization)1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation642481. Preventive maintenance 2. Calibration24216user1. Preventive maintenance 2. Calibration
EO injection stage(Fill the cabinet with EO gas)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Not enough ethylene oxide is entering the cabinetMissing EOThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1.EO cylinder hole alarm 2.EO injection timeout alarm10521001. Operator training 2. EO weight check before cycle start252201. Operator training 2. EO weight check before cycle startuser
Faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby period1.EO injection timeout alarm652601. Preventive maintenance652601. Preventive maintenanceuser
EO pipeline leakageBatches that exceed the qualification period may be sterilized againnone653901. Preventive maintenance 2. LEL EO detector alarm252201. Preventive maintenance 2. LEL EO detector alarmuser
Pressure sensor failureEO concentration is lower than the validation value (re-sterilization)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
2Liquid EO enters the cabinetLow carburetor temperatureThe cycle is locked in one phase until it is restored and restarted after a standby period1. The carburetor temperature is lower than the limit alarm642481. Preventive maintenance242161. Preventive maintenanceuser
Not enough water in the carburetor due to a faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby period1. The carburetor water level is lower than the limit alarm642481. Preventive maintenance242161. Preventive maintenanceuser
3Too much ethylene oxide input in the cabinetFaulty valveLonger quarantine, requiring additional testing1. Indoor high voltage alarm633541. Preventive maintenance 2. EO residual inspection233181. Preventive maintenance 2. EO residual inspectionuser
Pressure sensor failureLonger quarantine, requiring additional testing1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit632361. Preventive maintenance 2. Calibration232121. Preventive maintenance 2. Calibrationuser
4The temperature inside the cabinet exceeds the upper limitTemperature sensor failureImpact on product function (sorting)1. Redundant temperature sensor 2. Chamber high temperature alarm 3. Temperature difference exceeds limit alarm642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
5The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation632361. Preventive maintenance232121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Automatic heating time compensation633541. Preventive maintenance233181. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
Low carburetor temperatureAffects product sterility (re-sterilization)1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation642481. Preventive maintenance242161. Preventive maintenanceuser
Nitrogen cap stage(Add N2 to cover EO)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Not enough nitrogen entering the cabinetLack of nitrogenThe cycle is locked in one phase until it is restored and restarted after a standby period1. Check N2 pressure before starting the cycle 2. Alarm when N2 pressure is lower than the limit33218N/A
Valve failureThe cycle is locked in one phase until it is restored and restarted after a standby periodnone633541. Preventive maintenance233181. Preventive maintenanceuser
Pressure sensor failureFlammable Zone Hazard1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit632361. Preventive maintenance232121. Preventive maintenanceuser
2Too much nitrogen injected into the cabinetFaulty valveThe cabin is under positive pressure. This cycle will end.1. Indoor high voltage alarm 2. End of process6541201. Preventive maintenance254401. Preventive maintenanceuser
3The temperature inside the cabinet exceeds the upper limitTemperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Chamber high temperature alarm 3. Temperature difference exceeds limit alarm642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
4The box temperature is lower than the lower limit valueHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Circulating water pump overload alarm 3. Automatic heating time compensation632361. Preventive maintenance232121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Automatic heating time compensation633541. Preventive maintenance233181. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
Low carburetor temperatureAffects product sterility (re-sterilization)1. Alarm when the carburetor temperature is lower than the limit 2. Automatic heating time compensation642481. Preventive maintenance242161. Preventive maintenanceuser
EO exposure stage(Disinfect the product)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Lack of timeProgramming ErrorsRepeat sterilization1. Automatic cycle inspection252201. Preventive maintenance152101. Preventive maintenanceuser
2The box temperature is lower than the lower limit tHot water circulation pump failureThe cycle is locked at one stage until it is restored and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.Heating timeout 2. Circulating water pump overload alarm 3. Automatic heating time compensation632361. Preventive maintenance232121. Preventive maintenanceuser
Electric heating tube failureThe cycle is locked in one phase until it is resumed and restarted after a standby period. Batches that have exceeded the qualified cycle may be sterilized again.1. Heating timeout alarm 2. Automatic heating time compensation633541. Preventive maintenance233181. Preventive maintenanceuser
Temperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Temperature difference exceeds limit alarm652601. Preventive maintenance 2. Calibration252201. Preventive maintenance 2. Calibrationuser
3The temperature inside the cabinet exceeds the upper limitTemperature sensor failureAffects product sterility (re-sterilization)1. Redundant temperature sensor 2. Chamber high temperature alarm 3. Temperature alarm642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
4Not getting a sufficiently homogeneous EO mixtureCirculation fan failureAffects product sterility (re-sterilization)1. Low speed alarm653901. Preventive maintenance253301. Preventive maintenanceuser
Product positioning very close to the cavity wallAffects product sterility (re-sterilization)1.Guardrail is used to prevent the product from contacting the cavity wall253300N/A
EO Removal Phase(Create vacuum/remove sterilization gas)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Not purging enough gasInefficient vacuum pumpThe cycle is locked in one phase until it is restored and restarted after a standby period1.EO removal timeout alarm 2.Vacuum pump overload alarm 3.Automatically detect water flow and shut down the waterless pump632361. Preventive maintenance232121. Preventive maintenanceuser
Faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby period1.EO removal timeout alarm 2.Vacuum pump overload alarm632361. Preventive maintenance232121. Preventive maintenanceuser
Pressure sensor failureLonger isolation time and additional testing required due to less EO gas being removed1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit632361. Preventive maintenance 2. Calibration232121. Preventive maintenance 2. Calibrationuser
2Exhaust too much gasPressure sensor failureProduct and/or packaging damage (classification)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
Nitrogen purge stage(Ensure that the indoor environment is not flammable)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Not enough nitrogen entering the chamberLack of nitrogenThe cycle is locked in one phase until it is restored and restarted after a standby period1. Check N2 pressure before starting the cycle 2. Alarm when N2 pressure is lower than the limit33218N/A
Faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby periodnone633541. Preventive maintenance233181. Preventive maintenanceuser
Pressure sensor failureFlammable Zone Hazard1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit632361. Preventive maintenance232121. Preventive maintenanceuser
Longer quarantines require additional testing1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit632361. Preventive maintenance232121. Preventive maintenanceuser
Air cleaning stage(Create vacuum/remove residual sterilization gas, eliminate residual EO)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Not purging enough gasInefficient vacuum pumpThe cycle is locked in one phase until it is restored and restarted after a standby period1.EO removal timeout alarm 2.Vacuum pump overload alarm 3.Automatically detect water flow and shut down the waterless pump632361. Preventive maintenance232121. Preventive maintenanceuser
Faulty valveThe cycle is locked in one phase until it is restored and restarted after a standby period1.EO removal timeout alarm 2.Vacuum pump overload alarm632361. Preventive maintenance232121. Preventive maintenanceuser
Pressure sensor failureLonger isolation time and additional testing required due to less EO gas being removed1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit632361. Preventive maintenance 2. Calibration232121. Preventive maintenance 2. Calibrationuser
2Exhaust too much gasPressure sensor failureProduct and/or packaging damage (classification)1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit642481. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
3Insufficient air in the cabinFaulty valveThe cycle is locked in one phase until it is restored and restarted after a standby period1. Gas injection time over limit alarm643721. Preventive maintenance243241. Preventive maintenanceuser
Filter cloggedThe cycle is locked in one phase until it is restored and restarted after a standby period1. Gas injection time over limit alarm643721. Preventive maintenance243241. Preventive maintenanceuser
Pressure sensor failureLonger quarantines require additional testing1. Redundant pressure sensor 2. Alarm when pressure difference exceeds limit632361. Preventive maintenance 2. Calibration242161. Preventive maintenance 2. Calibrationuser
Removal stage(Remove residue from the product)
Serial numberPotentially risky contentRisk assessment under existing process conditionsRisk assessment of the results after taking measuresRecommended ActionsResponsibilities and current status
Fault nameCauseFailure ImpactControl NoweventSeverityexamineRisk PrioritizationTake actioneventSeverityexamineRisk Prioritization
1Aeration time is less than the specified timeProgramming ErrorsRepeat sterilization1. Automatic cycle inspection232121. Restart your computer regularly13261. Restart your computer regularlyuser