3.4.1 eto निर्जंतुकीकरण मशीन तापमान सेटिंग श्रेणी
The temperature setting range for the eto sterilizer is: 30 to 70°C, with a control error of ?2°C.

3.4.2 Temperature Uniformity of the ethylene oxide machine
The temperature uniformity of the eto sterilizer must meet the following requirements:

When the volume of the sterilization chamber is ?5m³, at least 10 measurement points must be placed.
When the volume is >5m³, for every additional 1m³, one more measurement point should be added.
When the volume is ?10m³, at least 20 measurement points must be set inside the chamber, with the test temperature ?40°C, and the temperature difference between each measurement point must not exceed ±3°C.
3.4.3 Sterilization Chamber Leakage Rate
3.4.3.1
Testing should be conducted before the injection phase of the sterilizing agent and the humidification phase (if using an inserted temperature and pressure display device). The pressure should be preset by the manufacturer, and the test pressure should be less than -20 kPa and at least 20 kPa lower than the preset sterilization pressure.

3.4.3.2
If the pressure during any stage of the sterilization cycle is below atmospheric pressure, after reaching the preset pressure, all valves connected to the sterilization chamber should be closed, and the vacuum pump should stop working. The pressure in the sterilization chamber should be monitored for at least 5 minutes, during which the pressure rise must not exceed 0.3 kPa/min.

3.4.3.3
When the working pressure of the eo chamber is greater than 5 kPa, a positive pressure leakage test should also be conducted, which should be performed before the vacuum stage. By pressurizing with inert gas (such as air) to within 10% of the preset working pressure, the pressure in the sterilization chamber should be monitored for at least 5 minutes, during which the pressure drop must not exceed 0.3 kPa/min.

3.4.4 eto sterilizer machine Vacuum Rate
Under no-load conditions, the eo chamber should reach -15 kPa from atmospheric pressure in ?6 minutes; reach -50 kPa in ?30 minutes; and reach -75 kPa in ?40 minutes.

3.4.5 Humidification Effect of the eto sterilizer machine
नो-लोड परिस्थितीत, जेव्हा सापेक्ष आर्द्रता ? 40% RH असते, तेव्हा इओ निर्जंतुकीकरण मशीनमध्ये आर्द्रता प्रणालीद्वारे सापेक्ष आर्द्रता किमान 75% किंवा त्याहून अधिक वाढवण्याची क्षमता असावी.

3.4.6 निर्जंतुकीकरण चेंबरची दाब सेटिंग श्रेणी
निर्जंतुकीकरण कक्षासाठी दाब सेटिंग श्रेणी -0.08 MPa ते +0.08 MPa आहे.

3.4.7 eto निर्जंतुकीकरण मशीनची कमाल आवाज पातळी
eto स्टेरिलायझेशन चेंबरद्वारे निर्माण होणारी कमाल आवाज पातळी (A-भारित) ?65 dB असावी.

3.4.8 इथिलीन ऑक्साईड निर्जंतुकीकरणाचे स्वरूप आणि रचना
३.४.८.१
आकार सरळ, गुळगुळीत आणि स्वच्छ पृष्ठभागासह, एकसमान रंग आणि बुरशी, तीक्ष्ण कडा आणि क्रॅक नसलेला असावा. कोणतेही स्पष्ट ओरखडे किंवा अडथळे यांसारखे दोष नसावेत.

३.४.८.२
पृष्ठभागावरील सर्व मजकूर, ग्राफिक्स, चिन्हे इत्यादी स्पष्ट, अचूक आणि टिकाऊ अशा खुणा सह स्पष्टपणे मुद्रित केले पाहिजेत.

३.४.८.३
फास्टनर्स सुरक्षितपणे स्थापित केले पाहिजेत आणि सर्व कंट्रोल स्विचेस आणि ऍडजस्टमेंट नॉब्स (बटणे) लवचिकपणे आणि विश्वासार्हपणे, कोणत्याही अडथळ्याशिवाय कार्य करावे.