5.1 Cov xwm txheej ua haujlwm li qub

5.1.1 Qhov chaw ua haujlwm ib txwm ua rau lub sterilizer yuav tsum ua raws li cov hauv qab no:

a) Ambient temperature: 5 ? to 40 ?;
b) Relative humidity: ? 80 % RH;
c) Atmospheric siab: 70 kPa rau 106 kPa;

5.1.2 Rated Kev Ua Haujlwm Ntawm Lub Sterilizer

Cov kev ntsuas kev ua haujlwm rau txhua tus qauv ntawm cov sterilizer muaj nyob rau hauv Table 2.

QauvCov khoom siv hluav taws xobNtau zausLub zog (kW)Tam sim no (A)
BCS-1380v ua50 Hz1624
BCS-21624
BCS-31624
BCS-4.52741
BCS-52741
BCS-62741
BCS-82741
BCS-104365
BCS-155685
BCS-2081123
BCS-2681123
BCS-2890137
BCS-3090137
BCS-30-290137
BCS-35100152
BCS-35-2100152
BCS-40-2100152
BCS-48100152
BCS-52150228
BCS-55-2150228
BCS-57-2150228
BCS-63-2150228
BCS-70-2155236
BCS-78-2165251
BCS-86-2165251
BCS-101-2200304
BCS-120-2200304
  • Lub qhov (socket) fais fab mov yog peb theem 380V, peb lub qhov (socket) nrog cov xaim hauv av.
  • Lub hwj chim plug yog peb-theem 380V, peb-prong plug nrog ib tug grounding hlau.
  • Ib lub xaim daj-ntsuab sab nraud yog siv los txuas rau qhov chaw tiv thaiv tus neeg xyuas pib.

5.2 Cia ib puag ncig ntawm 12m3 eto sterilizer

The sterilizer should be stored in an indoor environment with temperatures between -20? and +55?, relative humidity ? 80% RH, free from corrosive gases, and well-ventilated.
Nco tseg: Ua ntej cia, tag nrho cov dej nyob rau hauv lub circulating dej system yuav tsum tau drained kom lub system nyob rau hauv ib tug qhuav lub xeev.

5.3 Kev thauj mus los ntawm 12m3 eto sterilizer

Lub sterilizer yuav tsum tau thauj mus los tsis tau ntim, thiab thaum hoisting lub sterilizer, nws yuav tsum tau nqa nrog lub cuab yeej nqa lub nplhaib. Thaum thauj los ntawm lub tsheb, nws yuav tsum muaj kev ruaj ntseg raws li kev cai.

5.4 Tshem tawm cov khoom pov tseg 12m3 Sterilizers

Tom qab decommissioning, lub sterilizer yuav tsum muab pov tseg los ntawm lub tuam txhab tsim nyog recycling.