eto dauðhreinsunarferli

Implementation of the ETO Sterilization Process

Eto Sterilization is conducted using a eto sterilizer located in a dedicated facility. Outside the sterilization room, control equipment regulates the process according to established parameters and requirements.

Composition of Agents Used in the RÁÐU UPP Ófrjósemisaðgerð

The agents employed in sterilization consist of either pure ethylene oxide or mixtures of carbon dioxide and nitrogen.

Precautionary Measures for Ethylene Oxide

Ethylene oxide (EO) gas must be handled, stored, and used with extreme caution due to its highly toxic nature, flammability, and explosion risk at concentrations as low as 2.6% in air.

Steps Involved in the Sterilization Process

The eto sterilization process comprises two critical steps:

  1. Preconditioning
  2. Ófrjósemisaðgerð

Preconditioning

Preconditioning involves achieving the necessary humidity and temperature levels prior to introducing the load into the chamber, ensuring effective sterilization unaffected by internal environmental factors.

Sterilization Cycle

The EO sterilization process includes the following elements:

1. Upphitun
2. Haltu hita
3. Pre-tæmi
4.Lekaprófun
5.Rakagjöf
6.Rakabústaður
7.EO innspýting
8. EO innspýting2
9.EO innspýting 3
10.EO bústaður
11.Fjarlæging IT 2
12.Loftþvottur
13.Loftun

Influence of Load Chamber

The configuration of the load within the chamber affects:

  • Heat distribution patterns
  • Humidity levels
  • EO gas penetration
  • EO gas removal

Design of Load Configuration

The load configuration should be designed to facilitate the sterilization process by:

  • Removing air
  • Aiding heat penetration
  • Eliminating humidity
  • Ensuring EO removal during and after the process

Process Characterization & Safe Delivery of Sterilization Processes

Ferliseinkenni skilgreinir svið breytna sem krafist er samkvæmt ISO 11135 og auðkennir fjórar lykilferlisbreytur, sem þarf að skjalfesta á hverju stigi dauðhreinsunarferlisins:

  1. Áfangi
    Nær yfir breytur sem tengjast hverjum áfanga ferlisins.

  2. Forkæling
    Felur í sér tíma, hitastig, rakastig og flutningstíma.

  3. Sterilization Cycle
    Inniheldur útsetningartíma, hitastig, raka, EO styrk og þrýsting.

  4. Loftun
    Tilheyrir tíma og hitastigi meðan á loftun eða gasun stendur.

Skrunaðu efst