BOCON ETO STERILIZER
We offer: FAT, SAT, DQ, IQ, OQ, and support for PQ. Standard compliance: EN 1422, CE, ISO 13485।
Engineering Support
Custom dimensions or automatic conveyor loading systems available.
INQUIRE NOW →Industry ETO Sterilizer Turnkey EO Sterilization Project
A turnkey project is a complete project that is designed, built, and tested by a single provider. In the context of ETO sterilization, a turnkey project typically involves designing and building a complete ETO sterilization facility, including :
* preheating room
* Aeration Room
* Scrubber
* Sterilization conveyor system
* control system cabinet
* local control box
* etc
We are the main designated eto sterilizer supplier for Medline, Cardinal, and Medtronic. Medline, Cardinal, and Medtronic are all Fortune 100 companies, and the sterilization cabinets we provide account for over 60% of their total sterilization cabinet needs.
In addition to our partnerships with global leaders, we also serve as the main ethylene oxide sterilizer supplier for ALLMED, WINNER, and ZHENDE—the three largest medical device manufacturers in China. Our sterilization cabinet solutions account for over 90% of their sterilization equipment needs, further solidifying our position as a market leader.
BOCON Company has a strong R&D team, and we invest 2~3 million RMB in R&D expenses every year. We have 28 patents, including 3 Chinese invention patents and 2 American invention patents. Many companies focus their energy on sales, but we have always been dedicated to product research and development.
You could view our catalog to know more about our Chamber.
ETO Sterilizer Chamber Door System
Pneumatic Revolving Door
Pneumatic Sliding Door
Overall Lifting Door
- Automatic quality assessment report for software, which is an important function to assist operators and quality control personnel in quickly assessing production quality and identifying problems. Our sterilization cabinet software has more than 400 built-in automatic assessment functions, which can almost meet the quality automatic detection and assessment needs of all customers
- Very detailed software audit tracking function. Specifically, it is manifested by more than 120 automatic detection and prediction alarm functions, and almost all operations will be recorded in the system, making it easy to quickly identify the root cause of problems and repair them in the later stage.
- Standardized backend MES interface, with the customer’s consent, data can be automatically uploaded to cloud servers or the customer’s own backup server
- And more
Our control system has undergone long-term strict audits by Medline, Cardinal, Medtronic, WINNER, ZHENDE, and ALLMED companies, and has the most comprehensive functions and the best stability.
Why Choose BOCON EO Sterilizer?
(Equipped with Gas-water separator)
Based on China Industrial Power Rates
Water-Saving Technology
for Large-scale Facilities
1. Vacuum Pump Technology
| Item | BOCON | OTHERS |
|---|---|---|
| Vacuum Pump Options | Dry Screw / Water Ring / Oil Ring (Multiple options available) | Usually only Water Ring Pump |
| Ultimate Vacuum Level | Up to -101.27 kPa (Dry Screw Pump) | Approx. -90 kPa |
| Water Consumption | With BOCON Gas-water separator: Water replaced once per week | Continuous water consumption |
| Maintenance | Low maintenance (Dry pump), No oil emulsification risk | Frequent oil change (Oil pump) or limited performance |
2. Water-Saving Performance
Example: A 5.5 kW water ring pump consumes approximately 20 liters per minute. For a 20m³ sterilizer chamber with 8 vacuum cycles (22 minutes each), total water consumption per cycle reaches 3.52 tons.
With 14 cycles per week, total weekly consumption exceeds 49 tons. After installing BOCON’s exclusive Gas-water separator system, water replacement is required only once per week, saving approximately 45 tons of water per week (approx. 180 tons per month). This is especially valuable in water-scarce regions.
3. Exclusive EG Burner System
| Item | BOCON | OTHERS |
|---|---|---|
| Exhaust Treatment | Gas + Liquid waste combustion with energy recovery | Combustion only (gas) or scrubber only |
| Energy Recovery | High-temperature flue gas heats water for sterilizer use | No energy recovery |
| Liquid Waste Handling | Integrated liquid waste combustion | Scrubber cannot treat liquid waste |
The BOCON EG Burner combines waste liquid treatment and energy recycling. For example, a 100m³ sterilizer system can save approximately RMB 230,000 per year in electricity costs based on China industrial power rates. It also enables catalyst recovery, reducing ongoing operating expenses.
Note: Others currently mostly use combustion-only exhaust treatment, which involves higher risk and needs to stay on at all times, or use scrubbers which cannot handle liquid waste.
4. Centralized Heating System
| Item | BOCON | OTHERS |
|---|---|---|
| System Design | Centralized heating with dual pump redundancy | Single water heating unit |
| System Stability | Backup pump ensures continuous operation | System stops if pump fails |
| Application | Suitable for multiple sterilizers in one workshop | Individual heating per unit |
Our Partners










Safety of our EO sterilization equipment
EO is a flammable, explosive chemical. Its fire and explosion characteristics are system dependent. Some of the
characteristics for EO/air mixtures are as follows:
• Theminimumvaluecitedforthelowerflammable limit of EO air mixtures is 2.6% EO[20].
• The upper flammable limit is typically stated to be 100% because under certain conditions pure EO can violently
decompose with a significant release of heat in the absence of air or oxygen.
• Accordingly, the flammable range of EO-air mixtures is 2.6–100% EO.
• The autoignition temperature of EO in air at 14.7 psia is 833°F (445°C)
Bocon use mixed gas processes and full Nitride protection Processes to keep it safe.
Mixed gas processes
Full Nitride protection Processes
Frequently Asked Questions
Everything you need to know about BOCON ETO Sterilization technology.
Q1. What is the typical cycle time for an ETO sterilization process?
The entire ETO sterilization cycle usually takes between 12 to 24 hours. This includes critical stages: pre-conditioning (temperature/humidity adjustment), gas injection, exposure, and a lengthy aeration period to safely remove residual EO gas.
Q2. Why choose EO sterilization over Steam sterilization for medical devices?
EO is a low-temperature process (37°C to 55°C), making it ideal for heat-sensitive materials like plastics, electronics, and delicate surgical instruments that would be damaged by the high heat and moisture of steam autoclaves.
Q3. Are BOCON sterilizers compliant with ISO 11135 and EN 1422?
Yes. All BOCON equipment is manufactured to meet ISO 11135 and EN 1422 international standards. We provide full validation documentation support, including FAT, SAT, DQ, IQ, OQ, and assist with your PQ.
Q4. How do you ensure environmental safety regarding EO gas emissions?
BOCON ETO sterilizers can be integrated with our high-efficiency EO Scrubber system. This system uses chemical hydrolysis to neutralize 99.9% of ethylene oxide gas before it is exhausted, ensuring compliance with EPA and local environmental regulations.
Q5. What safety features are integrated into the Siemens PLC control system?
Our systems include explosion-proof components, real-time pressure and temperature monitoring, gas leak detection, and automatic interlocking doors. The Siemens PLC ensures precise control and records all cycle data for traceability.
Have more technical questions?
Contact our engineers for a free consultation →
















rebort –
great eto sterilizer,
१५६५७६१५९३५a –
Thank you very much for the high-quality products and services provided by Bocon. Welcome to visit my ETO sterilization workshop at any time.”
Glen –
The products and services are excellent!
I am happy to recommend these good EO sterilizers to my friends.
Welcome friends from Indonesia to consult me.
alex –
We purchased a 30m³ ETO sterilization system from BOCON for our new facility. Their engineering team provided exceptional support during the DQ, IQ, and OQ phases. The equipment fully complies with ISO 11135 standards, and we passed our audit without any issues. Highly professional!
CEO, Medical Packaging Solutions (Thailand) –
Excellent communication and after-sales service. Even with the time difference, BOCON’s technical support team responds quickly to our inquiries. The training they provided to our staff was thorough and very helpful.
Technical Engineer, Sterilization Services (India) –
We have been running BOCON ETO machines for over three years now. The uptime is impressive, and maintenance is straightforward. Compared to European brands, BOCON offers the best balance of advanced technology and cost-efficiency.” ——