1.3 30 m³ eto gas sterilizer Composition

The 30 m³ eto gas sterilizer consists of the sterilizer housing, auxiliary equipment (heating and thermal cycling device, vacuum extraction device, dosing and vaporization device, humidification device, sealing device, residual gas processing device), and monitoring and control device.

1.4 Classification

Based on the size of the eto gas sterilizer, there are two types: Class A and Class B.

  • Class A eto gas sterilizer — User-programmable sterilizer with a sterilization chamber volume > 1 m³, suitable for sterilizing in the medical device manufacturing industry.
  • Class Beto gas sterilizer — A sterilizer with one or more preset sterilization cycles and size limitations, with a sterilization chamber volume ? 1 m³, suitable for sterilizing clinical medical devices.

1.5 Model Difference Table

Table 1: Models and Basic Parameters of Ethylene Oxide Sterilizers

Serial No.แบบอย่างInternal Chamber Dimensions (L×W×H) (mm) ±1%Nominal Volume (m³)
กำลัง (กิโลวัตต์)Weight (kg)Power Supply
1บีซีเอส-11400*700*10001m³161000380V/50Hz
2บีซีเอส-21650*1000*12002m³162000380V/50Hz
3บีซีเอส-31400*1350*17003.0m³163000380V/50Hz
4บีซีเอส-4.52000*1350*17004.5m³274000380V/50Hz
5บีซีเอส-52250*1350*17005m³274500380V/50Hz
6บีซีเอส-62700*1350*17006 ลบ.ม275000380V/50Hz
7บีซีเอส-83500*1350*17008m³276000380V/50Hz
8บีซีเอส-104500*1350*170010ลบ.ม437000380V/50Hz
9บีซีเอส-155500*1350*200015m³569000380V/50Hz
10บีซีเอส-207500*1350*200020ลบ.ม8112000380V/50Hz
11บีซีเอส-269600*1350*200026m³8114000380V/50Hz
12บีซีเอส-2812500*1350*170028m³9014000380V/50Hz
13บีซีเอส-3011300*1350*200030ลบ.ม9017000380V/50Hz
14บีซีเอส-30-25600*2700*200030m³-29016000380V/50Hz
15บีซีเอส-3513000*1350*200035ลบ.ม10023000380V/50Hz
16บีซีเอส-35-26000*2700*220035m³-210022000380V/50Hz
17บีซีเอส-40-26650*2700*220040m³-210023000380V/50Hz
18บีซีเอส-4812850*1400*270048m³10028000380V/50Hz
19บีซีเอส-5213900*1400*270052m³15028000380V/50Hz
20บีซีเอส-55-27500*2700*270055m³-215027000380V/50Hz
21บีซีเอส-57-29650*2700*220057m³-215028000380V/50Hz
22บีซีเอส-63-28650*2700*270063m³-215029000380V/50Hz
23บีซีเอส-70-29700*2700*270070m³-215532000380V/50Hz
24บีซีเอส-78-210750*2700*270078m³-216535000380V/50Hz
25บีซีเอส-86-211800*2700*270086m³-216538000380V/50Hz
26บีซีเอส-101-217050*2700*2200101m³-220045000380V/50Hz
27บีซีเอส-120-220200*2700*2200120m³-220049000380V/50Hz

Sterilization Cycle and Automatic Control

2.10.7.1

The 30 m³ eto gas sterilizer should be equipped with an automatic controller that can automatically control the sterilization cycle.

2.10.7.2 Class A eto gas sterilizer:

  • The automatic controller should be able to program preset conditions for each stage of the sterilization cycle.
  • During the sterilization cycle, it must not change the preset variables.

2.10.7.3 Class B eto gas sterilizer:

  • The manufacturer applies the automatic controller to program the preset conditions for each stage of the sterilization cycle.
  • The manufacturer should provide methods to shorten the sterilization time for verification purposes, which can only be performed using a special key, password, or tool. Once the sterilization time is shortened, there should be a fault indication at the end of the operating cycle.

2.10.7.4

The controller should only be accessible using a special key, password, or tool. Additionally: a) For Class A sterilizers, access to the controller and setting of operational cycle variables should require a special key, password, or tool.
b) For Class B sterilizers, only the manufacturer can program new operational cycles.

2.10.7.5

During maintenance, testing, or emergencies, the program of the automatic controller should be operable manually. The manual device used should differ from the special key, password, or tool specified in section 2.10.7.4. When manual operation is performed due to maintenance, testing, or emergencies, the automatic program should stop automatically. The manual operation system should not pose a safety threat, allowing only one stage to be selected sequentially at a time.

2.10.7.6

Upon successful completion of each specified sterilization cycle, the automatic controller should indicate that the cycle is complete.