| Physical Performance Qualification (PPQ) • It shall be demonstrated that all specified acceptance criteria are reproducible throughout the entire process. How many runs should the PPQ process perform? • To demonstrate process reproducibility, at least 3 runs are required; • If 3 half-cycles were run during the MPQ process, then at least 1 additional run with full exposure time shall be performed. |
Reference Standards:
- ISO11135:2014 (Medical device EO sterilization validation)
- FDA Guidance (2007): Sterility Assurance for Terminal Sterilized Products
- EU GMP Annex 1 (2022): Sterilization process validation requirements.
How is Reproducibility Demonstrated?
| 唔得。 | Process Step | 項目 | 方法 |
|---|---|---|---|
| 1 | 預調 | Temperature and humidity of the sterilized load are within the specified range at the end of preconditioning | 測量負荷入面 |
| 2 | 載入緊 | 預調結束至開始殺菌之間嘅時間間隔應該喺指定範圍內 | 驗證記錄咗嘅時間數據 |
| 3 | eo 殺菌 | 一氧化乙烯( EO )會以氣體形式引入殺菌器 | 測量氣體注入溫度 |
| 4 | eo 殺菌 | 由於 EO 氣體注入而增加嘅壓力係喺指定範圍內 | 喺氣體注入期間測量壓力 |
| 5 | eo 殺菌 | 殺菌器入面嘅 EO 氣體濃度喺指定範圍內 | 喺注射前/後秤氣罐,然後計算差異或者 直接測量室內嘅氣體濃度 |
| 6 | eo 殺菌 | 喺曝光階段,負載入面嘅產品溫度喺指定範圍內 | 測量產品溫度 |
| 7 | 通風 | 喺通氣階段,負載入面嘅產品溫度喺指定範圍內 | 測量產品溫度 |
喺評估 PPQ 測試嘅執行情況嗰陣,應該注意以下幾個方面:
• 溫度同濕度探頭嘅數量係咪符合要求?
• (溫度同濕度)探頭係咪放喺裝咗嘅包裝入面,同埋放喺最壞情況嘅位置(最難殺菌嘅位置)?
• 產品負荷嘅溫度係咪喺指定範圍內?