喺 eo 殺菌嘅性能資格( pq ) - ppq

Physical Performance Qualification (PPQ)
• It shall be demonstrated that all specified acceptance criteria are reproducible throughout the entire process.
How many runs should the PPQ process perform?
• To demonstrate process reproducibility, at least 3 runs are required;
• If 3 half-cycles were run during the MPQ process, then at least 1 additional run with full exposure time shall be performed.

Reference Standards:

  • ISO11135:2014 (Medical device EO sterilization validation)
  • FDA Guidance (2007): Sterility Assurance for Terminal Sterilized Products
  • EU GMP Annex 1 (2022): Sterilization process validation requirements.

How is Reproducibility Demonstrated?

唔得。Process Step項目方法
1預調Temperature and humidity of the sterilized load are within the specified range at the end of preconditioning測量負荷入面
2載入緊預調結束至開始殺菌之間嘅時間間隔應該喺指定範圍內驗證記錄咗嘅時間數據
3eo 殺菌一氧化乙烯( EO )會以氣體形式引入殺菌器測量氣體注入溫度
4eo 殺菌由於 EO 氣體注入而增加嘅壓力係喺指定範圍內喺氣體注入期間測量壓力
5eo 殺菌殺菌器入面嘅 EO 氣體濃度喺指定範圍內喺注射前/後秤氣罐,然後計算差異或者 直接測量室內嘅氣體濃度
6eo 殺菌喺曝光階段,負載入面嘅產品溫度喺指定範圍內測量產品溫度
7通風喺通氣階段,負載入面嘅產品溫度喺指定範圍內測量產品溫度

喺評估 PPQ 測試嘅執行情況嗰陣,應該注意以下幾個方面:
• 溫度同濕度探頭嘅數量係咪符合要求?
• (溫度同濕度)探頭係咪放喺裝咗嘅包裝入面,同埋放喺最壞情況嘅位置(最難殺菌嘅位置)?
• 產品負荷嘅溫度係咪喺指定範圍內?