Prestazione di u Sterilizzatore ETO & Specificazioni | Guida tecnica chjave

Ethylene Oxide Sterilizer: Complete Performance Specifications & Compliance

Master the critical design, material, safety, and cycle requirements for ETO sterilizers. Based on industry standards, this guide covers everything from chamber construction to automatic control and leakage testing.

Informate via WhatsApp ? View ETO Sterilizer Product Details

Ethylene Oxide (EO/ETO) sterilization remains a cornerstone for heat-sensitive medical devices. To ensure safety, efficacy, and regulatory compliance, the sterilizer must meet rigorous performance indicators. Below we break down the key specifications – from materials and door interlocks to automatic cycle control and validation.

1. Construction, Materials & Dimensions

Apparizione & Struttura

L'esterno di u sterilizzatore deve esse liscia, uniforme in culore, senza bavature, crepe o graffi visibili. Tutti i fasteners, l'interruttori è i pomoli duveranu esse sicuri è operanu senza ostruzzioni.

Dimensioni di a camera

E camere rectangulari sò definite da a lunghezza, a larghezza, l'altitudine è u voluminu; camere cilindriche per diamitru, prufundità è voluminu. U spaziu utilizzabile deve mantene una distanza minima di 15 mm da superfici riscaldate è 30 mm da superfici senza riscaldamentu.

Cumpatibilità Materiale

Tutte e parti in cuntattu cù l'EO, u vapore, l'acqua o l'aria compressa devenu resistenti à a corrosione, micca reagisce cù l'EO, è ùn prumove micca a polimerizazione. A gomma naturale è u lattice sò pruibiti. Materiali appruvati cumuni: acciaio inox (? 8% nichel), ABS, PTFE, gomma di silicone.

2. Piping, Vaporizer & Fornitu

Piping & Cunnessioni

  • I tubi di sterile liquidu deve esse senza saldatura è senza perdite.
  • I giunti (eccettu a cunnessione di u cilindru EO) sò saldati.
  • I tubi chì portanu gas sterilizante o vapore necessitanu un insulamentu termale.
  • U tubu flessibile per EO deve aduprà acciaio inox, PTFE-lined, o gomma nitrile.
  • L'identificazione chjara di u mediu è a direzzione di u flussu hè ubligatoriu.

Vaporizatore & EO Supply

  • U vaporizzatore riscalda EO prima di l'entrata di a camera; a superficia di riscaldamentu deve esse pulita.
  • A temperatura di u gasu EO dopu a vaporizzazione ùn deve micca più di 70 ° C.
  • U riscaldamentu direttu di cartucce o cilindri EO hè pruibitu.
  • Un filtru (?50 ?m) hè stallatu in a linea di supply EO à u vaporizzatore.
  • Dosage basatu à u pesu: precisione di scala in ± 5% di a carica EO.
3. Sistemi di porta & Interlocks di sicurezza

Sigillatura di a porta & Accessu

Doors must have replaceable seals, cleanable without removing the door. Before cycle start, loading door can be opened. Once cycle begins, door remains locked until “cycle complete” – unlocking requires a dedicated key, code, or tool.

Automatic & Manual Doors

Automatic doors stay locked and sealed if power fails. Manual door operation force must not exceed 150 N. Radial locking arms need a removable guard for inspection.

Double-Door Sterilizers

Simultaneous opening of loading and unloading doors is prevented except for maintenance. Indicators on both sides (and remote control room) show which door can be opened.

4. Instrumentation & Control automaticu

Mandatory instruments include:

  • Chamber pressure indicator
  • Chamber temperature indicator
  • Sterilization phase indicator
  • Cycle complete indicator
  • Door status indicator
  • Jacket temperature indicator
  • EO inlet temperature (if injection >30s)
  • Temperature & pressure recorders
  • Leak rate pressure monitor

Temperature & Precisione di pressione

Temperature : range 0–100°C, accuracy ±1%, resolution 1°C, indication error <±1°C.
Pressure: range 0 kPa to max working pressure, accuracy ±1.6%, sterilization pressure error <±5 kPa. Leak test monitor: 20 kPa span, ±1% accuracy, 0.1 kPa resolution.

Recording & Data Integrity

Permanent records required (no thermal paper that fades). Sampling interval ?2 seconds. Digital records must be tamper-proof. Analog charts: speed ?1 mm/min, width ?100 mm. Digital records print cycle start/end values with out-of-tolerance alerts.

Automatic Controller

Class A sterilizers allow user programming with key/code; Class B are factory-programmed only. Over-temperature protection independent of control system. Software must be validated. Faults trigger visual/audible alarm and safe state.

5. Sterilization Cycle Stages (9 Phases)
  1. Chamber pre-heating
  2. Vacuum phase
  3. Automatic leak test
  4. Conditioning (heat & humidify)
  5. EO injection
  6. Sterilization exposure
  7. EO exhaust
  8. Aeration / washing
  9. Air admission & cycle end

Key requirements: Leak test must pass (pressure change ?0.3 kPa/min). Conditioning for Class B must maintain relative humidity >40%. Sterilization timing starts only when temperature, pressure, and concentration reach setpoints. After cycle, if door isn’t opened immediately, additional aeration or continuous flushing must ensure safe EO levels.

6. Performance, Electrical Safety & EMC
  • Chamber surface temperature ? setpoint +5°C.
  • Empty chamber temperature uniformity: ±3°C of setpoint.
  • Vacuum system must achieve -75 kPa or lower.
  • Leak rate: ?0.1 kPa/min for volumes ?1 m³.
  • Noise level ?65 dB(A).
  • Asbestos-free insulation; surfaces >55°C must be insulated.
  • Biological indicator inactivation required for Class B sterilizers.
  • Electrical safety per GB4793.1 / IEC 61010-2-040.
  • Electromagnetic compatibility per GB/T18268.1.
  • Sterilant verification independent of injection control (pressure, weight, or volume).
  • Fault diagnosis measures for maintenance.