Author name: Bocon Technical Team

Eto গ্যাস জীবাণুমুক্তকারী প্রস্তুতকারকের জন্য অনুরোধ

2.1.1 The overall shape should be precise, with smooth, clean surfaces and uniform color. There should be no burrs, sharp edges, cracks, visible scratches, or unevenness.
2.1.2 All external text, graphics, and symbols must be printed clearly, with precise and durable markings.
2.1.3 Fasteners must be securely installed. All control switches and adjustment knobs (or buttons) should operate smoothly and reliably without obstructions.

Eto গ্যাস জীবাণুমুক্তকারী প্রস্তুতকারকের জন্য অনুরোধ আরও পড়ুন

ETO Sterilization Chamber Testing

The temperature setting range of the ETO Sterilization Chamber should be between 30-60°C. It is recommended to perform temperature distribution testing of the sterilizer’s internal surface and temperature and humidity distribution testing inside the sterilization chamber to ensure the effectiveness of the heating and humidifying systems.

ETO Sterilization Chamber Testing আরও পড়ুন

Routine Monitoring, Control, and Release of Ethylene Oxide Sterilization

The purpose of routine monitoring and control of EO (Ethylene Oxide) sterilization is to demonstrate that the confirmed and specified sterilization process has been applied to the product. Data from each sterilization cycle should be recorded and retained to prove that the confirmed sterilization process specifications have been met. This should include at least the following content:

Routine Monitoring, Control, and Release of Ethylene Oxide Sterilization আরও পড়ুন

ইও নির্বীজন প্রক্রিয়া চ্যালেঞ্জ ডিভাইস উত্পাদন

A Process Challenge Device (PCD) is a microbiological challenge system used to evaluate the kill rate of selected process parameters. It is typically a device or test package containing biological indicators (e.g., Bacillus atrophaeus spores with known spore count). The resistance of the PCD to the sterilization process should be greater than or equal to the resistance of the natural microbial load in the most difficult-to-sterilize area of the product. The quantity of PCDs should meet the requirements outlined in Table C.3 of Appendix C in GB18279.1-2015. PCDs are classified into Internal PCDs (IPCD) and External PCDs (EPCD).

ইও নির্বীজন প্রক্রিয়া চ্যালেঞ্জ ডিভাইস উত্পাদন আরও পড়ুন

ইন্ডাস্ট্রি ইটো স্টেরিলাইজার ইনস্টলেশন যাচাইকরণ

Industry Eto Sterilizer Installation Verification The installation of Industry Eto Sterilizer and all associated services should comply with the requirements

ইন্ডাস্ট্রি ইটো স্টেরিলাইজার ইনস্টলেশন যাচাইকরণ আরও পড়ুন

ETO জীবাণুমুক্তকরণের জন্য বিশুদ্ধ জলের প্রয়োজনীয়তা

The humidification water used in the ETO Sterilization Equipment should be free from contaminants, should not weaken the ETO sterilization process, and should not damage the sterilizer, sterilized products, or the load. Humidification should be done by injecting steam, not by directly injecting atomized water.

ETO জীবাণুমুক্তকরণের জন্য বিশুদ্ধ জলের প্রয়োজনীয়তা আরও পড়ুন

ETO Sterilizer Manufacturer-HZBOCON

Hangzhou Bocon Mechanical and Electrical Equipment Co., Ltd., commonly referred to as HZBOCON, stands out as a dedicated Ethylene Oxide (EO/ETO) sterilizer manufacturer. Established in 2006, HZBOCON has committed itself to the design, development, and manufacturing of EO sterilization equipment, catering to a wide array of industries that require effective sterilization solutions for thermosensitive products.

ETO Sterilizer Manufacturer-HZBOCON আরও পড়ুন

ইথিলিন অক্সাইড নির্বীজন চেম্বারে PQ

PQ হল বৈধতার পর্যায় যা ইথিলিন অক্সাইড নির্বীজন চেম্বার ব্যবহার করে দেখায় যে ইথিলিন অক্সাইড জীবাণুমুক্তকরণ সরঞ্জামগুলি ধারাবাহিকভাবে পূর্বনির্ধারিত গ্রহণযোগ্যতার মানদণ্ড অনুসারে কাজ করে এবং প্রক্রিয়াটি এমন একটি পণ্য তৈরি করে যা জীবাণুমুক্ত এবং নির্দিষ্ট প্রয়োজনীয়তা পূরণ করে।

ইথিলিন অক্সাইড নির্বীজন চেম্বারে PQ আরও পড়ুন